Skip to the content

FDA Medical Device Recalls (openFDA)

Tripath Imaging, Inc.: PrepStain GN 480 Disposable Kit Catalog/Model 490542 This is a complete kit, per the firm's explanation, that contains the PrepStain Syringing Pipettes 240, the product being recalled. The affected lots are listed under those pipettes (PrepStain Syringing Pipettes 240)

Recall number
Z-1815-2012
Recalling firm
Tripath Imaging, Inc.
Product
PrepStain GN 480 Disposable Kit Catalog/Model 490542 This is a complete kit, per the firm's explanation, that contains the PrepStain Syringing Pipettes 240, the product being recalled. The affected lots are listed under those pipettes (PrepStain Syringing Pipettes 240)
Status
Terminated
Initiated
2012-03-22
Posted
2012-06-15
Terminated
2012-12-03
Reason
The PrepStain Syringing Pipettes may not dispense the minimum volume of 7.5 mL of the sample due to leaking. This event resulted in an increase in complaints of leaking syringing pipettes
Root cause
Component design/selection
Action
BD Diagnostics sent a "BD PrepMate Syringing Pipette" letter dated March 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The firm encouraged customers to refer to the Operator's Manual and inspect the centrifuge tubes to verify that the correct amount of sample has been transferred. Customers experiencing this issue were instructed to discard the pipettes and contact BD Diagnostic Technical Support for a replacement. Contact the firm at 1-877-822-7771 for questions regarding this notice.
Quantity
46,112,784 for all products in Recall Event
Distribution
Worldwide Distribution.
Lot, serial or model codes
This is a complete kit, per the firm's explanation, that contains the PrepStain Syringing Pipettes 240, the actual product being recalled. The affected lots are listed under those pipettes (PrepStain Syringing Pipettes 240), which are: 100901a, 100902f, 100902f, 100903d, 100908c, 100910c, 100914f, 100916d, 100921d, 100928c, 101005a, 101007d, 101008c, 101012c, 101013a, 101015c, 101020d, 101022c, 101027a, 101028c, 101029d, 101103b, 101105h, 101110f, 101111c, 101117p, 101118n,  101119c, 101130p, 101201o, 101207o, 101208p, 101210c, 101213n, 101214b, 101214c,  101217b, 101220q, 101221g, 110104b, 110106a, 110107c, 110110d, 110112n, 110118b,  110128i, 110203i, 110210r
510(k) numbers
not on file
PMA numbers
["P970018"]
Product code
MKQ
Firm FEI
3008007472
Firm city
Burlington
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register