Skip to the content

FDA Medical Device Recalls (openFDA)

Hospira Inc: Hospira Phoenix Infusion System with MedNet Software, Symbiq One-Channel Infusion System, Software versions v2.1 & v3.0, Manufactured by Hospira Inc., Morgan Hill, CA.

Recall number
Z-1815-2009
Recalling firm
Hospira Inc
Product
Hospira Phoenix Infusion System with MedNet Software, Symbiq One-Channel Infusion System, Software versions v2.1 & v3.0, Manufactured by Hospira Inc., Morgan Hill, CA.
Status
Terminated
Initiated
2009-03-31
Posted
2009-08-11
Terminated
2010-12-18
Reason
Potential delay/underinfusion of critical therapy-- Devices experience increased frequency of alarms that result in temporary pump stoppage.
Root cause
Software design
Action
Hospira, Inc. issued an "Urgent: Device Field Correction" dated March 31, 2009 via Federal Express to Healthcare Professionals and Valued Hospira Customers. Consignees were informed of the affected devices and will be contacted by a Hospira representative to arrange necessary device upgrades. For further questions, contact Hospira Global Product Safety and Complaints at 1-800-441-4100.
Quantity
not on file
Distribution
Product was distributed to 59 consignees to all 50 States.
Lot, serial or model codes
List number 16026-04-83/84 and 87/88, software versions 2.1 and 3.0 only.
510(k) numbers
["K041550"]
PMA numbers
not on file
Product code
frn
Firm FEI
2921482
Firm city
Morgan Hill
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register