Ev3, Inc: ev3 IntraCoil Self-Expanding Peripheral Stent, Sterile EO, Model REF VT-5-60-135, Lot 4401432, Use Before 2010-07. The IntraCoil is indicated for use in peripheral arteries in the leg in which balloon angioplasty has been applied. The IntaCoil stent is indicated for placement in flexing regions and a tortuous lesion.

Recall number
Z-1815-2008
Recalling firm
Ev3, Inc
Product
ev3 IntraCoil Self-Expanding Peripheral Stent, Sterile EO, Model REF VT-5-60-135, Lot 4401432, Use Before 2010-07. The IntraCoil is indicated for use in peripheral arteries in the leg in which balloon angioplasty has been applied. The IntaCoil stent is indicated for placement in flexing regions and a tortuous lesion.
Status
Terminated
Initiated
2008-04-02
Posted
2008-09-11
Terminated
2011-12-10
Reason
Incorrect Label - an IntraCoil 5mm x 40 mm stent was labeled as a 5 mm x 60 mm stent (VT-5-40-135 was labeled as VT-5-60-135).
Root cause
Employee error
Action
ev3 representative retrieved from the consignnee the affected device on 4/4/08. A Device Reconcilliation Form was requested to be filled and returned with the product.
Quantity
3
Distribution
Nationwide Distribution including states of CA and MO.
Lot, serial or model codes
Lot No: 4401432
510(k) numbers
not on file
PMA numbers
["P000033"]
Product code
NIP
Firm FEI
2183870
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register