FDA Medical Device Recalls (openFDA)
Aesculap Implant Systems LLC: AESCULAP MINOP InVent 30 Trocar System, non-sterile Product Usage:The Minop InVent Trocar System intended use is for endoscopic procedures within the central nervous system, especially for the treatment of intra- and paraventricular pathological structures.
- Recall number
- Z-1814-2017
- Recalling firm
- Aesculap Implant Systems LLC
- Product
- AESCULAP MINOP InVent 30 Trocar System, non-sterile Product Usage:The Minop InVent Trocar System intended use is for endoscopic procedures within the central nervous system, especially for the treatment of intra- and paraventricular pathological structures.
- Status
- Terminated
- Initiated
- 2017-03-07
- Posted
- 2017-03-20
- Terminated
- 2018-02-16
- Reason
- Aesculap Implant Systems LLC is recalling the Minop Trocar due to the possibility it may have sharp edges on the distal end which may lead to the abrasion of the insulation when removing the electrode.
- Root cause
- Nonconforming Material/Component
- Action
- On March 15, 2017, 16 facilities and 1 Sales Rep were sent an Urgent Medical Device Recall Notification letter. Letters were sent Fed-Ex overnight. Customers were asked to immediately discontinue use and quarantine the product. A Sales Representative will remove the affected product and return to Aesculap Inc.
- Quantity
- 21 units distributed in U.S.
- Distribution
- Product was distributed throughout the United States and Canada.
- Lot, serial or model codes
- Item # FH620R
- 510(k) numbers
- ["K983365"]
- PMA numbers
- not on file
- Product code
- GWG
- Firm FEI
- 2916714
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28