Skip to the content

FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES, LP - Northfield: Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ACCESSORY PACK SKU DYNJ59270 2. CRANI PACK-LF SKU DYNJ44805M 3. CRANIOTOMY SKU CDS982719V CDS982719W DYNJ904168F DYNJ908404B DYNJ908723A DYNJ908723C 4. CRANIOTOMY #IMF 56646-LF SKU DYNJVB91001A 5. CRANIOTOMY CDS SKU CDS981888X CDS983467I 6. CRANIOTOMY PACK-LF SKU DYNJ09882G DYNJ09882I 7. CRANIOTOMY S NASSAU SKU DYNJ905879G 8. MHC CRANIOTOMY SKU DYNJ903713K DYNJ903713L 9. PAD CRANIOTOMY SKU DYNJ902149J

Recall number
Z-1811-2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ACCESSORY PACK SKU DYNJ59270 2. CRANI PACK-LF SKU DYNJ44805M 3. CRANIOTOMY SKU CDS982719V CDS982719W DYNJ904168F DYNJ908404B DYNJ908723A DYNJ908723C 4. CRANIOTOMY #IMF 56646-LF SKU DYNJVB91001A 5. CRANIOTOMY CDS SKU CDS981888X CDS983467I 6. CRANIOTOMY PACK-LF SKU DYNJ09882G DYNJ09882I 7. CRANIOTOMY S NASSAU SKU DYNJ905879G 8. MHC CRANIOTOMY SKU DYNJ903713K DYNJ903713L 9. PAD CRANIOTOMY SKU DYNJ902149J
Status
Open, Classified
Initiated
2025-04-22
Posted
2025-05-31
Terminated
not on file
Reason
Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.
Root cause
Nonconforming Material/Component
Action
On April 22, 2025, the firm notified customers of the recall through letters titled Immediate Action Required Voluntary Medical Device Recall. Customers were instructed to identify and quarantine all affected product on hand. Customers are to submit a response form to the recalling firm. Upon receipt of the response firm, Medline will provide the customer with over-labels to place on affected inventory, which will include instructions for staff to remove and discard the affected component prior to using the kit. Distributors should notify downstream customers/accounts of this recall. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
1,853 total
Distribution
US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.
Lot, serial or model codes
1. CRANI ACCESSORY PACK SKU DYNJ59270 UDI-DI 10888277771512 (ea), 40888277771513 (case) Lots 23JDB683 23JDC548 24ADB783 24HDA822 2. CRANI PACK-LF SKU DYNJ44805M UDI-DI 10195327506872 (ea), 40195327506873 (case) Lots 23JBI747 23KBL228 24ABG696 24CBM117 24EBQ114 24GBJ699 3. CRANIOTOMY SKU CDS982719V UDI-DI 10195327510831 (ea), 40195327510832 (case) Lots 23KBP461 23LBN381 24ABB119 24BBC547 24BBI145 SKU CDS982719W UDI-DI 10195327645090 (ea), 40195327645091 (case) Lots 24DBS813 SKU DYNJ904168F UDI-DI 10195327395728 (ea), 40195327395729 (case) Lots 24DBU188 SKU DYNJ908404B UDI-DI 10195327410506 (ea), 40195327410507 (case) Lots 23IDB680 23LDB546 24BDB174 24FDA741 24FDB584 24IDA225 24IDA646 24IDA840 SKU DYNJ908723A UDI-DI 10195327403270 (ea), 40195327403271 (case) Lots 23IBO314 23JBC218 24ABK090 SKU DYNJ908723C UDI-DI 10198459009785 (ea), 40198459009786 (case) Lots 24IBM205 24IBN365 25ABM356 4. CRANIOTOMY #IMF 56646-LF SKU DYNJVB91001A UDI-DI 10195327466466 (ea), 40195327466467 (case) Lots 23JLA765 24DLA724 24DLB006 24ELA325 5. CRANIOTOMY CDS SKU CDS981888X UDI-DI 10195327144760 (ea), 40195327144761 (case) Lots 24ABT432 24CBD698 24HBH954 24HBP483 24IBG797 25CBF098 SKU CDS983467I UDI-DI 10195327457945 (ea), 40195327457946 (case) Lots 23KDA336 6. CRANIOTOMY PACK-LF SKU DYNJ09882G UDI-DI 10195327574055 (ea), 40195327574056 (case) Lots 24CBF778 SKU DYNJ09882I UDI-DI 10198459058172 (ea), 40198459058173 (case) Lots 24HBN426 7. CRANIOTOMY S NASSAU SKU DYNJ905879G UDI-DI 10195327512996 (ea), 40195327512997 (case) Lots 23KBI086 23LBH647 23LBH649 24BBU253 24DBV133 24FBE409 8. MHC CRANIOTOMY SKU DYNJ903713K UDI-DI 10195327201579 (ea), 40195327201570 (case) Lots 23JBK426 SKU DYNJ903713L UDI-DI 10195327531218 (ea), 40195327531219 (case) Lots 23KBT939 23LBS862 24ABH577 24CBH126 24CBH771 9. PAD CRANIOTOMY SKU DYNJ902149J UDI-DI 10195327554323 (ea), 40195327554324 (case) Lots 23KBP275 24ABJ684 24CBF216 24DBI296 24EBD959 24FBF064 24HBK164 24IBN443
510(k) numbers
not on file
PMA numbers
not on file
Product code
OJH
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register