OptumHealth Care Solutions LLC: Optum Nimbus II Plus, Model number AAA-00170; Ambulatory Infusion Pump

Recall number
Z-1811-2024
Recalling firm
OptumHealth Care Solutions LLC
Product
Optum Nimbus II Plus, Model number AAA-00170; Ambulatory Infusion Pump
Status
Open, Classified
Initiated
2024-04-18
Posted
2024-05-21
Terminated
not on file
Reason
Nimbus II Plus, Infusion Pump Systems, manufactured by InfuTronix, are being removed the from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump housing.
Root cause
Device Design
Action
Optumhealth Care Solutions, LLC issued an Urgent Device Recall by the Manufacturer of Nimbus II Pump notice to its consignees on 04/18/2024 via email Delivery/Adobe Sign and UPS Next Day Air. The notice sent an electronic link to the press release and stated that a nurse would be contacting the consignee to schedule a pump replacement, including a home visit to instruct the patient on the use of the new pump. In the meantime, if you have any issue with your current Nimbus II pump, please contact your nurse right away to prioritize your pump replacement. Additionally, the manufacturer recommends in its recall notice that you take the following actions, so please be aware and comply with them until your Nimbus II pump has been replaced: 1. When replacing batteries, only new batteries should be used. Care should be taken to ensure that old and new batteries are not confused when completing the replacement. 2. The battery should not be cycled (turned on/off) to clear the occlusion alarm. If the occlusion alarm cannot be cleared, a new pump should be used. Silencing the alarm does not clear the occlusion. 3. Due to the potential ambulatory use of the pump, you may carry the drug product dispensing pouch in a carrying pack. Please note that impeding the tubing set or placing pressure on the device (leaning against it, sitting on it) may cause the device to malfunction. 4. If you note any difficulties or abnormalities with the pump or accessories, immediately discontinue use of the pump and notify your Optum nurse or health care provider.
Quantity
208 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
UDI/DI 00862492000308, Each unit has a Lot Number and a Serial Number, listed as follows: Lot Number/Serial Number: 302604 / NIMBP702231, 318050 / NIMBP715656, 340086 / NIMBS718475, 340163 / NIMBS716926, 340648 / NIMBS716122, 340746 / NIMBS715845, 340820 / NIMBS712929, 340903 / NIMBP719704, 341046 / NIMBS713994, 341372 / NIMBS718513, 341535 / NIMBS718700, 341757 / NIMBS716661, 341770 / NIMBS716996, 342136 / NIMBS715836, 342233 / NIMBS718503, 342350 / NIMBS714995, 342447 / NIMBS713444, 342938 / NIMBS716864, 343066 / NIMBS716125, 343309 / NIMBS719492, 343683 / NIMBS716506, 344023 / NIMBS716532, 344044 / NIMBS718477, 344173 / NIMBS718500, 344273 / NIMBS715037, 344407 / NIMBS709122, 344440 / NIMBS709111, 344450 / NIMBS708973, 344455 / NIMBS709568, 344572 / NIMBS709989, 344588 / NIMBS708676, 344692 / NIMBS709563, 344783 / NIMBS709112, 344853 / NIMBS709482, 344862 / NIMBS709556, 344976 / NIMBS709267, 345057 / NIMBS716989, 345060 / NIMBS716509, 345190 / NIMBS708707, 345195 / NIMBS719668, 345396 / NIMBS718671, 345397 / NIMBS708966, 345443 / NIMBS716367, 345446 / NIMBS719686, 345728 / NIMBS718488, 345732 / NIMBS716946, 345774 / NIMBS716166, 345872 / NIMBS709020, 345879 / NIMBS716159, 346002 / NIMBS716398, 346156 / NIMBS710038, 346167 / NIMBS709114, 346318 / NIMBS709337, 346410 / NIMBS715276, 346634 / NIMBS709569, 346636 / NIMBS709080, 346638 / NIMBS717005, 346642 / NIMBS716878, 346643 / NIMBS709550, 346947 / NIMBS709316, 347056 / NIMBS718701, 347173 / NIMBS716827, 347296 / NIMBS717651, 347304 / NIMBS708964, 347396 / NIMBS713609, 347526 / NIMBS716794, 347529 / NIMBS716803, 347534 / NIMBS718487, 347650 / NIMBS716650, 347654 / NIMBS716906, 347664 / NIMBS716952, 347665 / NIMBS716487, 347666 / NIMBS718494, 347799 / NIMBS716814, 347827 / NIMBS717661, 347828 / NIMBS718490, 347854 / NIMBS709506, 347856 / NIMBS706539, 347858 / NIMBS708541, 348023 / NIMBS715840, 348029 / NIMBS716933, 348032 / NIMBS717419, 348044 / NIMBS716161, 348150 / NIMBS716489, 348152 / NIMBS716507, 348159 / NIMBS716395, 348161 / NIMBS718657, 348251 / NIMBS715850, 348257 / NIMBS719132, 348358 / NIMBS709103, 348360 / NIMBS716999, 348447 / NIMBP719452, 348453 / NIMBS712917, 348535 / NIMBS717650, 348538 / NIMBS719666, 348598 / NIMBS716649, 348599 / NIMBS718672, 348605 / NIMBP719122, 348698 / NIMBS719665, 348704 / NIMBS716958, 348768 / NIMBS716929, 348770 / NIMBS717845, 348771 / NIMBS718687, 348822 / NIMBS709562, 348925 / NIMBS717642, 348926 / NIMBS716945, 348928 / NIMBS718685, 349012 / NIMBS709268, 349016 / NIMBS708438, 349043 / NIMBS708901, 349050 / NIMBS708930, 349076 / NIMBS708771, 349080 / NIMBS708617, 349131 / NIMBS709233, 349132 / NIMBS708612, 349147 / NIMBS708812, 349148 / NIMBS709155, 349151 / NIMBS709143, 349195 / NIMBS708753, 349242 / NIMBS709958, 349247 / NIMBS715726, 349281 / NIMBS716660, 349283 / NIMBS708623, 349329 / NIMBS719707, 349337 / NIMBS715730, 349338 / NIMBS716951, 349339 / NIMBS707669, 349429 / NIMBP707727, 349434 / NIMBS708976, 349454 / NIMBS719462, 349509 / NIMBS709794, 349562 / NIMBS718501, 349565 / NIMBS718699, 349566 / NIMBS719692, 349613 / NIMBS709124, 349707 / NIMBS708968, 349785 / NIMBS718666, 349786 / NIMBS718684, 349787 / NIMBS708998, 349911 / NIMBS714990, 349912 / NIMBS708068, 349914 / NIMBS718509, 350073 / NIMBS716807, 350077 / NIMBS707032, 350081 / NIMBS709119, 350083 / NIMBS708884, 350086 / NIMBS716950, 350087 / NIMBS716179, 350174 / NIMBS718682, 350175 / NIMBP703889, 350179 / NIMBS718695, 350208 / NIMBS709025, 350209 / NIMBS708937, 350211 / NIMBS709475, 350214 / NIMBS718667, 350216 / NIMBS719472, 350217 / NIMBS718677, 350270 / NIMBP715002, 350387 / NIMBS708659, 350388 / NIMBS719695, 350389 / NIMBS709065, 350516 / NIMBS718673, 350605 / NIMBS708437, 350698 / NIMBS708198, 350834 / NIMBS708807, 350835 / NIMBS708609, 350837 / NIMBS708916, 350838 / NIMBS709097, 350843 / NIMBS708750, 350845 / NIMBS716153, 350846 / NIMBS716150, 350847 / NIMBS716871, 350849 / NIMBS716986, 350851 / NIMBS713751, 350852 / NIMBS715994, 350854 / NIMBS708816, 350855 / NIMBS713878, 350856 / NIMBS716833, 350857 / NIMBS715015, 350860 / NIMBP719679, 350867 / NIMBS718492, 350868 / NIMBS716164, 350870 / NIMBS714987, 350886 / NIMBS715741, 350888 / NIMBS717664, 350889 / NIMBS709244, 350890 / NIMBS709436, 350891 / NIMBS718680, 350892 / NIMBS708427, 350893 / NIMBS708667, 350894 / NIMBS717652, 350895 / NIMBS718498, 350896 / NIMBS716949, 350897 / NIMBS717658, 350899 / NIMBS716914, 351697 / NIMBS719678, 351699 / NIMBS708735, 351701 / NIMBS716931, 351702 / NIMBS709326, 351703 / NIMBS718473, 351704 / NIMBS716400, 351715 / NIMBS708604, 351717 / NIMBS709269, 351720 / NIMBS709148, 351722 / NIMBS709473, 351723 / NIMBP716638, 351725 / NIMBS709060, 351898 / NIMBS718689.
510(k) numbers
["K153193"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3009456923
Firm city
Marietta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register