Skip to the content

FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES, LP - Northfield: Medline Kits containing Codman Disposable Perforator 14 mm: CRANIOTOMY, SKU DYNJ35349C

Recall number
Z-1810-2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline Kits containing Codman Disposable Perforator 14 mm: CRANIOTOMY, SKU DYNJ35349C
Status
Open, Classified
Initiated
2025-04-22
Posted
2025-05-31
Terminated
not on file
Reason
Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.
Root cause
Nonconforming Material/Component
Action
On April 22, 2025, the firm notified customers of the recall through letters titled Immediate Action Required Voluntary Medical Device Recall. Customers were instructed to identify and quarantine all affected product on hand. Customers are to submit a response form to the recalling firm. Upon receipt of the response firm, Medline will provide the customer with over-labels to place on affected inventory, which will include instructions for staff to remove and discard the affected component prior to using the kit. Distributors should notify downstream customers/accounts of this recall. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
1853 total
Distribution
US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.
Lot, serial or model codes
UDI-DI 10888277858060 (EA); 40888277858061 (CASE) KIT LOTS 24CBJ925 24DBT405 24FBO318
510(k) numbers
["K213481"]
PMA numbers
not on file
Product code
OJG
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register