FDA Medical Device Recalls (openFDA)
WELCH ALLYN, INC/MORTARA: Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.
- Recall number
- Z-1810-2020
- Recalling firm
- WELCH ALLYN, INC/MORTARA
- Product
- Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.
- Status
- Terminated
- Initiated
- 2020-03-18
- Posted
- 2020-04-29
- Terminated
- 2026-02-09
- Reason
- Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Root cause
- Device Design
- Action
- Welch Allyn notified customers on 2/19/2020 via "Voluntary Medical Device Correction Welch Allyn Patient Cables or Lead Sets". The recall letter identified the affected product and stated the reason for the recall. The firm also requested the distributors to immediately notify accounts if the product was further distributed.
- Quantity
- 78,521 units total
- Distribution
- US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
- Lot, serial or model codes
- Model Numbers: ELI10-XXX-XXXXX, ELI350-XXX-XXXXX, ELI150C-XXX-XXXXX, ELI250C-XXX-XXXXX, ELI230-XXX-XXXXX, ELI280-XXX-XXXXX, ELI380-XYYZZ, RSCRIBE-xxx-xxxxx and CC-XXX-XXXX.
- 510(k) numbers
- ["K092663","K100127","K101115","K101403","K102206"]
- PMA numbers
- not on file
- Product code
- DPS
- Firm FEI
- 2183461
- Firm city
- Milwaukee
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28