Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Instruments Div. of Stryker Corporation: Navigation System II-Cart, 7700-100-000 with Articulated Arm Camera, 7700-103-001 and Articulated Arm Camera, 7700-103-001U Product Usage: The Navigation System II-Cart is intended to be used as a platform for computer assisted surgery as part of the Stryker Navigation System. The intended use of the complete Stryker Navigation System is defined by the Stryker application modules.

Recall number
Z-1810-2015
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Product
Navigation System II-Cart, 7700-100-000 with Articulated Arm Camera, 7700-103-001 and Articulated Arm Camera, 7700-103-001U Product Usage: The Navigation System II-Cart is intended to be used as a platform for computer assisted surgery as part of the Stryker Navigation System. The intended use of the complete Stryker Navigation System is defined by the Stryker application modules.
Status
Terminated
Initiated
2015-05-04
Posted
2015-06-16
Terminated
2016-05-16
Reason
Stryker initiated a device recall of the Navigation System II-Cart, Camera Articulated Arm because there is a potential that the weld seam between the main stud and the two flaps on the Articulated Arm was not welded in the correct location, which may result in a potential failure of the arm joint.
Root cause
Process control
Action
Stryker sent an Urgent Medical Device Recall Notification letter dated May 18, 2013 to Domestic customers notice via FedEx overnight and Sales Representatives via email. The International Subsidiaries will also be notified via email. Customers are instructed to immediately review the Recall Notification adn complete the following: Immediately check all stock areas and/or operating room storage for affected equipment found. Complete the enclosed Business Reply Form (BRF) to confirm receipt of this notification and identify how many affected items are currently in inventory. Please complete and return the BRF even if they dont have any affected product on hand. If customers have further distributed this product, they are asked to forward this letter and the attached BRF to all affected locations. Please indicate each location and serial number(s) on the BRF. Fax (866-521-2762) or email (kellyjo.davis@stryker.com) the completed BRF to Stryker Instruments Regulatory Department, Attn: Kelly Jo Davis. A Stryker Representative will contact the facility to set up a time to perform the additional inspection of the weld seam on the Navigation System II-Cart, Articulated Arm Camera.
Quantity
72
Distribution
Worldwide Distribution - US Nationwide in the states of AR, CO, FL, GA, IN, KS, KY, MA, MI, NC, NE, NJ, NY, OH, OK, PA, TX, WA, WY and the countries of Canada, Europe, and Asia.
Lot, serial or model codes
Serial numbers 100019, 100984, 100965, 100295, 100979, 100773, 100946, 100797, 100947, 100956, 100966, 100649, 100967, 100944, 100941, 100945, 100970, 100950, 100976, 100951, 100728, 100952, 100248, 100953, 100641, 100954, 100463, 100955, 100981, 100957, 100978, 100958, 100106, 100959, 100939, 100960, 100973, 100961, 100236, 100962, 100237, 100963, 100417, 100964. 100940, 100968, 100975, 100969, 100977, 100972, 100105, 100974, 100214, 100980, 100492, 100982, 100493, 100983, 100948, 100986, 100704, 100987, 100949, 100988, 100227, 100989, 100971, 100133, 100134, 100942.
510(k) numbers
["K022365"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1811755
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register