FDA Medical Device Recalls (openFDA)
Cardiac Science Corporation: Cardiac Science Powerheart 9390A automated external defibrillator. This fully automatic model does not require the user to press a shock button in order for the device to deliver therapy (when appropriate). Intended for use in either in-hospital or out-of-hospital settings for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing.
- Recall number
- Z-1810-2010
- Recalling firm
- Cardiac Science Corporation
- Product
- Cardiac Science Powerheart 9390A automated external defibrillator. This fully automatic model does not require the user to press a shock button in order for the device to deliver therapy (when appropriate). Intended for use in either in-hospital or out-of-hospital settings for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing.
- Status
- Terminated
- Initiated
- 2009-12-18
- Posted
- 2010-06-21
- Terminated
- 2011-12-08
- Reason
- The K302 relay switch of the G3 series AED product line may experience an early failure. If the K302 contacts are open at the time of rescue, there is a possible delay in therapy or failure to deliver therapy.
- Root cause
- Other
- Action
- Customer notification letters were mailed to US end users and distributors (final consignees) between December 18 and 23, 2009 by USPS certified mail, return receipt requested. Notification letters were sent by FedEx to Canadian distributors (final consignees) on December 23, 2009. Secondary consignee (distributor) letters were mailed on December 23, 2009. Letters advise user to check the status indicator on front of AED to determine if it is rescue ready. Users are advised to keep the AED in service and to perform the Daily and Monthly Scheduled Maintenance as outlined in the Operator and Service Manual. Questions about this communication should be addressed to 1.877.901.1788 toll free from inside the US, or to email .
- Quantity
- 332 units
- Distribution
- Domestic: 323 units to distributors and end users. Canada: 9 units to 2 distributors.
- Lot, serial or model codes
- Serial Numbers: 4259985, 4260065, 4260137, 4260148, 4260174, 4260179, 4260181, 4261110, 4261111, 4261112, 4261113, 4261114, 4261115, 4261116, 4261117, 4261118, 4261119, 4261120, 4261121, 4261122, 4261123, 4261125, 4261126, 4261127, 4261128, 4261129, 4261130, 4261131, 4261132, 4261133, 4261134, 4261136, 4261138, 4261139, 4261140, 4261141, 4261142, 4261143, 4261144, 4261145, 4261147, 4261149, 4261150, 4261151, 4261153, 4261154, 4261155, 4261156, 4261157, 4261158, 4261160, 4261161, 4261162, 4261163, 4261164, 4261166, 4261168, 4261169, 4261170, 4261172, 4261174, 4261175, 4261177, 4261178, 4261179, 4261180, 4261181, 4261182, 4261183, 4261184, 4261185, 4261186, 4261188, 4261189, 4261190, 4261216, 4261217, 4261284, 4261285, 4261292, 4261299, 4261300, 4261301, 4261305, 4261306, 4261308, 4261312, 4261313, 4261317, 4261319, 4261320, 4261321, 4261325, 4261326, 4261329, 4261330, 4261332, 4261333, 4261342, 4261343, 4261348, 4261351, 4261352, 4261355, 4261356, 4261358, 4261360, 4261361, 4261365, 4261370, 4261381, 4261382, 4261383, 4261384, 4261385, 4261386, 4261387, 4261388, 4261394, 4261395, 4261396, 4261397, 4261398, 4261399, 4261400, 4261401, 4261402, 4261403, 4261404, 4261405, 4261406, 4261407, 4261408, 4261409, 4261410, 4261411, 4261412, 4261413, 4261414, 4261415, 4261416, 4261417, 4261420, 4261422, 4261423, 4261425, 4261449, 4261450, 4261451, 4261463, 4261465, 4261473, 4261475, 4261476, 4261477, 4261478, 4261479, 4261480, 4261481, 4261482, 4261483, 4261484, 4261485, 4261486, 4261487, 4261488, 4261489, 4261490, 4261492, 4261495, 4261496, 4261498, 4261499, 4261500, 4261501, 4261502, 4261503, 4261505, 4261506, 4261507, 4261508, 4261512, 4261513, 4261514, 4261515, 4261516, 4261517, 4261518, 4261519, 4261520, 4261522, 4261523, 4261524, 4261526, 4261527, 4261528, 4261529, 4261530, 4261531, 4261532, 4261534, 4261535, 4261536, 4261537, 4261538, 4261539, 4261540, 4261541, 4261542, 4261543, 4261544, 4261545, 4261546, 4261547, 4261548, 4261549, 4261551, 4261552, 4261554, 4261555, 4261557, 4261558, 4261559, 4261560, 4261561, 4261562, 4261563, 4261564, 4261565, 4261566, 4261567, 4261568, 4261570, 4261571, 4261572, 4261573, 4261574, 4261575, 4261576, 4261577, 4261579, 4261580, 4261586, 4261587, 4261592, 4261597, 4261598, 4261605, 4261608, 4261610, 4261611, 4261613, 4261614, 4261615, 4261617, 4261618, 4261621, 4261623, 4261624, 4261626, 4261628, 4261630, 4261631, 4261632, 4261633, 4261634, 4261635, 4261636, 4261637, 4261638, 4261639, 4261640, 4261641, 4261642, 4261643, 4261644, 4261645, 4261646, 4261647, 4261648, 4261649, 4261650, 4261651, 4261652, 4261653, 4261654, 4261656, 4261657, 4261660, 4261662, 4261663, 4261664, 4261665, 4261666, 4261670, 4261672, 4261676, 4261678, 4261679, 4261680, 4261683, 4261688, 4261689, 4261690, 4261691, 4261692, 4261695, 4261696, 4261697, 4261700, 4261709, 4261711, 4261712, 4261713, 4261714, 4261716, 4261718, 4261721, 4261726, 4261727, 4261729, 4261733, 4261734, 4261737, 4261739, 4261744, 4261745, 4261746, 4261747, 4261748, 4261750, AND 4261751.
- 510(k) numbers
- ["K031987"]
- PMA numbers
- not on file
- Product code
- MKJ
- Firm FEI
- 2112020
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28