Skip to the content

FDA Medical Device Recalls (openFDA)

DePuy Mitek, Inc., a Johnson & Johnson Co.: Lupine BR Anchor w/Orthocord TCP/PLGA Absorbable Anchor, size 2 (5 metric) Braided Composite Suture, violet 36" (91cm) Size 2 (5metric) eyelet loop. Made in Switzerland; Sterile. The Lupine BR Anchor System is indicated for use in soft tissue bone fixation.

Recall number
Z-1810-2008
Recalling firm
DePuy Mitek, Inc., a Johnson & Johnson Co.
Product
Lupine BR Anchor w/Orthocord TCP/PLGA Absorbable Anchor, size 2 (5 metric) Braided Composite Suture, violet 36" (91cm) Size 2 (5metric) eyelet loop. Made in Switzerland; Sterile. The Lupine BR Anchor System is indicated for use in soft tissue bone fixation.
Status
Terminated
Initiated
2008-04-29
Posted
2008-09-17
Terminated
2008-10-02
Reason
Lupine BR Anchor w/Orthocord was assembled with an incorrect suture configuration. The suture loop was not assembled at the distal end of the anchor. It was only threaded through the suture hole in the anchor. Risk associated with incorrect suture configuration include premature breakage.
Root cause
Employee error
Action
DePuy Mitek notified customers by letter dated April 29, 2008 titled Urgent: Voluntary Medical Device Recall Letter asking consignees to isolate and do not use product and to return any of the identified product to DePuy Mitek using enclosed prepaid Federal Express shipping labels and to include a photocopy of the completed Business Reply Form. The firm is asking customers to complete the Business Reply Form with a self addressed, stamped envelope indicating level of inventory of the affected product and to return for immediate delivery of replacement product. If you have any questions contact DePuy Mitek Customer Service directly at 1-800-382-4682.
Quantity
105 units
Distribution
Class II Recall - Worldwide Distribution --- USA including states of CA, NY, MA, NV, OH, TX, TN, MD, and countries of France, UK, Canada, Spain.
Lot, serial or model codes
Catalog Number 210711 Lot Number 3069512
510(k) numbers
["K070925"]
PMA numbers
not on file
Product code
JDR
Firm FEI
1221934
Firm city
Raynham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register