Philips North America, LLC: Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system Interventional fluoroscopic x-ray system System code: 722079

Recall number
Z-1809-2020
Recalling firm
Philips North America, LLC
Product
Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system Interventional fluoroscopic x-ray system System code: 722079
Status
Terminated
Initiated
2020-04-03
Posted
2020-04-29
Terminated
2023-02-01
Reason
When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.
Root cause
Software design
Action
Philips Healthcare issued Urgent Medical Device Correction letter via Field Action 2020-IGTBST-003 issued on April 3, 2020. Letter states reason for recall, health risk an action to take: To avoid this hazardous situation, the users should: BY CUSTOMER/ USER " Use XperGuide live guidance with the L-arm positioned in the XperCT scan position. " Move the L-arm back to the initial XperCT scan position if the system displays the warning Live 3D roadmap not possible. Please move stand to the : [acquisition position] position. At any stage during needle guidance, the user can perform a verification run to verify if the needle position is correct. Please ensure that all staff working with the XperGuide application are informed of the content of this letter and place a copy of this letter as an addendum to the Instructions for Use. Questions contact your local Philips representative. Technical Support Line: 1-800-722-9377. Philips will resolve the problem for all affected systems by installing a new Interventional Workspot software release. This action will start in April 2020. Expanded Recall 12-22-20: The two systems for the Netherland have already been corrected through the implementation of software update. For the system in Germany, the Customer has been notified through the Field Safety Notice.
Quantity
22 units (6 US and 16 OUS); Expanded 3units; Total: 25
Distribution
Worldwide distribution. US states of FL, ID, IL, IN, MA, and OH, Canada, Australia, Austria, Belgium, Germany, and Netherlands New Zealand, South Africa, Switzerland, Thailand
Lot, serial or model codes
Software:Philips Azurion 2.0 Interventional Fluoroscopic X-ray system with Interventional Workspot 1.5  US Serial Numbers: 637, 200, 909, 1276, 1309, 987  Canada: 1016   ROW: 11169 1129 1143 971 1135 824 934 1089 1008 1165 1080 1019 1248 523 1172  Expanded Recall OUS: 1223, 1330, 703424
510(k) numbers
["K181830"]
PMA numbers
not on file
Product code
OWB
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register