Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: iGUIDE System Product Usage: The intended use of the device is the control of accurate patient positioning with the assistance of a 30 Tracking System in a radiotherapy environment.

Recall number
Z-1807-2014
Recalling firm
Elekta, Inc.
Product
iGUIDE System Product Usage: The intended use of the device is the control of accurate patient positioning with the assistance of a 30 Tracking System in a radiotherapy environment.
Status
Terminated
Initiated
2014-05-29
Posted
2014-06-17
Terminated
2017-09-29
Reason
A bug in the software prevents the iGUIDE software from logging off.
Root cause
Software design
Action
Elekta sent an Important Field Safety Notice 618 01 303 024 dated May 29, 2014 to affected customers. The letter identified the affected product, problem, potential clinical impact and actions to be taken. It also gives a recommendation how the risk can be mitigated - Users should only use the iGUIDE software when the key is in the on position and the Position Guard is active. Users are advised to restart the workstation if any error occurs to ensure the software is running properly. The notice also advises the customer that a patch will be released that corrects the behavior of the iGUIDE software. This solution will be implemented on all affected products when available. For questions contact your local Elekta office.
Quantity
224
Distribution
Worldwide Distribution - USA (nationwide) in the states of AZ, CA, CT, HI, IL, IA, KY, MD, MA, MI, MO, NH, NJ, NC, ND, OH, OK, OR, PA, TX, VA, WA, WI, and the countries of Algeria, Australia, Austria, Brazil, Canada, China, Colombia, Denmark, Finland, France, Germany, Hong Kong, India, Italy, Japan, Kazakhstan, Mexico, Netherlands, New Zealand, Peru, Poland, Russia, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom and Vietnam
Lot, serial or model codes
Software Versions 2.0.0, 2.0.1 and 2.0.2
510(k) numbers
["K072079"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register