Skip to the content

FDA Medical Device Recalls (openFDA)

Acumed LLC: Ulnar Shorting Saw Blade is labeled in part: "Congruent Bone Plates, SIZE Ulnar Shortening Saw Blade - Hub Style S, REF 80-0499-S, Material: Stainless Steel, STERILE, Rx Only, ACUMED, Hillsboro, OR, USA". The Ulnar Saw Blade is an accessory used with a Congruent Bone Plate System covered under K012655, but it is labeled and sold separately by itself. The saw blade is also sold as part of the Upper Extremity Osteotomy System (BOM 15-0030). The BOM 15-0030 system does not have a label, but the saw blade is individually labeled and shipped together with this system when it is ordered. Intended use: Cutting Bones

Recall number
Z-1807-2010
Recalling firm
Acumed LLC
Product
Ulnar Shorting Saw Blade is labeled in part: "Congruent Bone Plates, SIZE Ulnar Shortening Saw Blade - Hub Style S, REF 80-0499-S, Material: Stainless Steel, STERILE, Rx Only, ACUMED, Hillsboro, OR, USA". The Ulnar Saw Blade is an accessory used with a Congruent Bone Plate System covered under K012655, but it is labeled and sold separately by itself. The saw blade is also sold as part of the Upper Extremity Osteotomy System (BOM 15-0030). The BOM 15-0030 system does not have a label, but the saw blade is individually labeled and shipped together with this system when it is ordered. Intended use: Cutting Bones
Status
Terminated
Initiated
2010-04-14
Posted
2010-06-16
Terminated
2010-06-28
Reason
The saw blades may not cut aggressively enough for their intended application therefore creating a risk that the saw could generate excessive heat, which may contribute to bone necrosis and/or non-union.
Root cause
Other
Action
On April 14, 2010, Acumed began contacting consignees by telephone. On April 15, 2010, Acumed sent out the URGENT NOTICE: DEVICE RECALL Initial Notification letter and Recalled Product Status Form to their consignees. The letter describes the recall product as Ulnar Shorting Saw Blade-Hub Style S, Part no. 80-0499-S and Lot Codes. The Initial Notification letter instructs consignees to return the product back to Acumed immediately and notify their sub-accounts (customers) if products have been further distributed. The consignees also are asked to fill out and send back the Recalled Product Status Form . Customers should call the Customer Service at 1-888-627-9957 or e-mail to RecallNotification@acumed.net for any questions regarding the recall.
Quantity
718 units
Distribution
Worldwide distribution: USA and to the following countries: Spain, Great Britain, South Africa, Puerto Rico, Canada, Turkey, Ireland, Finland, Belgium, Australia, Germany, and Sweden.
Lot, serial or model codes
Part Number: 80-0499-S Lot Numbers: 214436, 215637, 215638, 215639, 215640, 215641, 219931, and 220909.
510(k) numbers
["K012655"]
PMA numbers
not on file
Product code
HRS
Firm FEI
1000125930
Firm city
Hillsboro
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register