Skip to the content

FDA Medical Device Recalls (openFDA)

Obalon Therapeutics Inc: Obalon Balloon System with Model 4300 Touch Dispenser Touchscreen; Software version 02.00.00.0076 - Product Usage: The System is intended to be used as an adjunct to a moderate intensity diet and behavior modification program. All balloons must be removed 6 months after the first balloon is placed. The Obalon Balloons are administered using the Navigation System.

Recall number
Z-1805-2020
Recalling firm
Obalon Therapeutics Inc
Product
Obalon Balloon System with Model 4300 Touch Dispenser Touchscreen; Software version 02.00.00.0076 - Product Usage: The System is intended to be used as an adjunct to a moderate intensity diet and behavior modification program. All balloons must be removed 6 months after the first balloon is placed. The Obalon Balloons are administered using the Navigation System.
Status
Completed
Initiated
2019-05-23
Posted
not on file
Terminated
not on file
Reason
During initial set-up and/or replacement of the dispenser batteries, the touch dispenser touchscreen can inadvertently lose calibration. This issue can also occur if the touchscreen is pressed when the device is powered on in preparation for a balloon administration. As a result, the touchscreen can become unresponsive and balloon inflation cannot be initiated.
Root cause
Software design
Action
To ensure that customers would receive the recall notification bulletin, and to ensure that the bulletin would not be lost in an email inbox, customers were contacted by telephone beginning on 5/23/2019 to notify them that they would be receiving an email containing the notification bulletin. These emails, first sent on 5/23/2019, contained Customer Bulletin no. LIT-7410-0025-01. The bulletin notified customers of the unintended functionality of the touchscreen during power-up, provided instructions to contact Obalon Customer Service immediately it this situation occurs, and contained updated instructions as to how to handle the device issue. Customers were instructed to contact the Obalon Customer Service/Technical Support department to obtain instructions for re-calibrating the touchscreen such that the device will function as designed and the balloon administration may proceed. Customers were requested to immediately report any events that occur with use of the system such that the firm can inform regulatory bodies and customers of any new device information and update labeling materials, where appropriate. Customers were informed that a permanent fix has been identified and once it has been reviewed and approved by FDA, all dispensers will be updated accordingly.
Quantity
45 units
Distribution
US Nationwide distribution including the states of VA, NY, GA, TX, TN, OR, FL, LA, CT, and NJ.
Lot, serial or model codes
ObalonR Balloon System with Navigation and Touch. UDI: 00859810006104; Model No.: Model 4300 Touch Dispenser;Software version 02.00.00.0076. Serial Numbers: 0097, 0076, 0077, 0078, 0079, 0080, 0082, 0083, 0084, 0085, 0086, 0087 0088, 0089, 0090, 0091, 0092, 0093, 0094, 0095, 0097, 0099, 0100, 0101, 0102, 0103, 0104, 0106, 0107, 0109, 0111, 0112, 0113, 0115, 0118, 0119, 0120, 0121, 0122, 0125, 0126, 0127, 0129, 0130, 0133.
510(k) numbers
not on file
PMA numbers
["P160001S020","P160001S007"]
Product code
LTI
Firm FEI
3013508647
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register