Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes USA (HQ), Inc.: Synthes Medullary Tubes

Recall number
Z-1804-2010
Recalling firm
Synthes USA (HQ), Inc.
Product
Synthes Medullary Tubes
Status
Terminated
Initiated
2010-04-01
Posted
2010-06-16
Terminated
2015-03-25
Reason
After repeated sterilizations product may become discolored and break including while in the medullary canal. Synthes has determined that the Medullary Tube will be a single use item that is to be discarded after one sterilization cycle.
Root cause
Other
Action
The recalling firm issued an Urgent medical Device Recall letter dated 4/9/10 to all trauma hospital accounts informing them of the problem and the need to return the products. All sales consultants were notified via email and letter. Questions are directed to the company at 800-620-7025 x 5452 or 610-719-5452, or contact your Synthes Trauma Sales Consultant.
Quantity
18,505 tubes
Distribution
The product was shipped to medical facilities nationwide.
Lot, serial or model codes
Catalog number 355.01. all lot numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3005180112
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register