Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: EZPass Passer, Sterile single use arthroscopic instrument utilized in conjunction with a nitinol wire to aid in passing suture through soft tissue. Part numbers 904050, 904051, 904052, 904053, 904054, 904055, and 904056.

Recall number
Z-1803-2014
Recalling firm
Biomet, Inc.
Product
EZPass Passer, Sterile single use arthroscopic instrument utilized in conjunction with a nitinol wire to aid in passing suture through soft tissue. Part numbers 904050, 904051, 904052, 904053, 904054, 904055, and 904056.
Status
Terminated
Initiated
2014-04-15
Posted
2014-06-16
Terminated
2015-07-07
Reason
The nitinol wire or monofilament may buckle within the EZPass Suture Passer and become kinked making the wire/monofilament difficult to use.
Root cause
Device Design
Action
On 2/28/14, customers received a letter stating these devices have been recalled following an investigation which identified that the nitinol wire or monofilament may buckle within the EzPass Suture Passer and become kinked making the wire/monofilament difficult to use. The letter also stated, there is no adverse health outcome expected for the patient and this action requires the immediate location and discontinued use of the product and its return to Biomet.
Quantity
102
Distribution
Nationwide Distribution.
Lot, serial or model codes
Part # 904050 Lots M224900, M175380, M060840,Part ##904051 Lots M224930, M224970, M175390 and M060850,Part # 904052 Lots M298610, M225000, M225010, and M060860; Part #904053 Lots M225030, M225020 and M048380; Part # 904054 Lots M225040, M225050, M175420, M060870; Part # 904055 Lot M060880; and Part # 904056 Lots M225100, M175440, and M060890.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HWQ
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register