Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Peripheral Vascular Inc: BARD PERIPHERAL VASCULAR SEEKER(R) Crossing Support Catheter, 0.035", 150 cm, REF SK15035M. Intended to support a guidewire during access into the vasculature.

Recall number
Z-1802-2018
Recalling firm
Bard Peripheral Vascular Inc
Product
BARD PERIPHERAL VASCULAR SEEKER(R) Crossing Support Catheter, 0.035", 150 cm, REF SK15035M. Intended to support a guidewire during access into the vasculature.
Status
Terminated
Initiated
2018-04-16
Posted
2018-05-11
Terminated
2021-03-02
Reason
Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endotoxins are present on the packaging hoops it may have the potential for cross-contamination in a clinical setting.
Root cause
Process control
Action
The firm initiated the recall via an Urgent Medical Device Recall Notification letter on 04/16/2018. The letter identified the affected device, reason for the recall, and required actions. Customers are asked to check their inventory for and remove the affected product code and lot number combinations. The Recall and Effectiveness Check Form should be completed and returned even if affected product has been used. The firm requested the return of the recalled product; and will provide a replacement after all information has been verified and product returned. The completed form can be emailed to BPV.CustomerSupportCenter@crbard.com or faxed to 1-800-994-6772.
Quantity
140 units
Distribution
Nationwide Distribution
Lot, serial or model codes
UDI: (01)00801741067921(17)191115(10)VTBY0563, Lot Number: VTBY0563
510(k) numbers
["K103367"]
PMA numbers
not on file
Product code
DQY
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.