Philips North America, LLC: Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures.
- Recall number
- Z-1801-2020
- Recalling firm
- Philips North America, LLC
- Product
- Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures.
- Status
- Terminated
- Initiated
- 2020-03-05
- Posted
- not on file
- Terminated
- 2022-09-09
- Reason
- No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws becoming loose..
- Root cause
- Component design/selection
- Action
- The firm, Philips, sent an "URGENT-MEDICAL DEVICE CORRECTION" letter to the US and an "URGENT-Field Safety Notice" letter to the OUS (foreign countries) dated 2/26/20 on 3/5/2020 to all consignees by certified mail. These letters to all users explained that no torque was specified for the screws that connect the gearbox and the mounting flange within the C-arm roll motor assembly. A low torque could cause stress on the system and result in a loosening of the screws. If this occurs, the C-arm could rotate freely and uncontrolled movement could cause collision and/or entrapment. Philips asked that all users of the system be informed of the issues and to place a copy of the Field Safety Notice in the Instructions for Use for the system. Philips will exchange the roll motor in all affected systems free of charge, starting in March 2020. A local Philips representative will contact each consignee to schedule the roll motor replacement. If you need any further information or support concerning this issue, please contact your local Philips representative. Technical Support Line: 1-800-722-9377.
- Quantity
- 162 units
- Distribution
- Worldwide distribution: US (nationwide) including the following states: Arkansas, Arizona, California, Colorado, Delaware, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oregon, Pennsylvania, Puerto Rico, Texas, Virginia, Wisconsin. And the following Countries of: Australia, Azerbaijan, Chile, Croatia, Czechoslovakia, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Korea South, Lebanon, Mexico, Netherlands, Norway, Pakistan, Poland, Russian Federate, Senegal, Spain, South Africa, Sweden, Switzerland, Taiwan, and Uruguay.
- Lot, serial or model codes
- Serial Numbers: 709, 710, 711, 713, 716, 715, 439, 526, 637, 653, 684, 621, 497, 792, 595, 674, 737, 628, 429, 605, 895, 540, 769, 668, 697, 758, 633, 520, 738, 691, 813, 428, 590, 706, 611, 655, 804, 663, 588, 877, 854, 654, 488, 836, 689, 717, 790, 629, 598, 736, 810, 841, 946, 795, 673, 776, 645, 646, 703, 608, 805, 746, 262, 730, 753, 878, 755, 837, 618, 594, 714, 543, 751, 263, 766, 817, 741, 534, 535, 536, 466 System Equipment Numbers: 74620447, 74620437, 74613625, 74613538, 74620341, 77839007, 73340583, 73326701, 77646555, 77559977, 77970948, 77427180, 72722066, 79132604, 72273457, 76100048, 77970800, 74620302, 72805859, 77228623, 80018072, 74611961, 77647014, 75486269, 78248928, 75486357, 76888124, 75581571, 78243622, 75423992, 77971015, 74817549, 76890243, 74352012, 77260210, 77213972, 78993356, 77347585, 75532686, 79152378, 79109118, 75533239, 74536213, 79152254, 77838972, 78150020, 78696158, 77559984, 76481840, 75848181, 79112403, 77905831, 80171865, 79042710, 77697096, 77213970, 74778339, 74778897, 73051857, 73717291, 78132484, 75305793, 72312081, 76890941, 76888973, 78365267, 76888978, 79464469, 72323525, 72323935, 74169336, 74620425, 76891009, 72417922, 79035318, 79035262, 78367631, 74584486, 74584491, 75615010, 75470091
- 510(k) numbers
- ["K181830"]
- PMA numbers
- not on file
- Product code
- JAA
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28