Philips North America, LLC: Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures.

Recall number
Z-1801-2020
Recalling firm
Philips North America, LLC
Product
Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures.
Status
Terminated
Initiated
2020-03-05
Posted
not on file
Terminated
2022-09-09
Reason
No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws becoming loose..
Root cause
Component design/selection
Action
The firm, Philips, sent an "URGENT-MEDICAL DEVICE CORRECTION" letter to the US and an "URGENT-Field Safety Notice" letter to the OUS (foreign countries) dated 2/26/20 on 3/5/2020 to all consignees by certified mail. These letters to all users explained that no torque was specified for the screws that connect the gearbox and the mounting flange within the C-arm roll motor assembly. A low torque could cause stress on the system and result in a loosening of the screws. If this occurs, the C-arm could rotate freely and uncontrolled movement could cause collision and/or entrapment. Philips asked that all users of the system be informed of the issues and to place a copy of the Field Safety Notice in the Instructions for Use for the system. Philips will exchange the roll motor in all affected systems free of charge, starting in March 2020. A local Philips representative will contact each consignee to schedule the roll motor replacement. If you need any further information or support concerning this issue, please contact your local Philips representative. Technical Support Line: 1-800-722-9377.
Quantity
162 units
Distribution
Worldwide distribution: US (nationwide) including the following states: Arkansas, Arizona, California, Colorado, Delaware, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oregon, Pennsylvania, Puerto Rico, Texas, Virginia, Wisconsin. And the following Countries of: Australia, Azerbaijan, Chile, Croatia, Czechoslovakia, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Korea South, Lebanon, Mexico, Netherlands, Norway, Pakistan, Poland, Russian Federate, Senegal, Spain, South Africa, Sweden, Switzerland, Taiwan, and Uruguay.
Lot, serial or model codes
Serial Numbers:  709, 710, 711, 713, 716, 715, 439, 526, 637, 653, 684, 621, 497, 792, 595, 674, 737, 628, 429, 605, 895, 540, 769, 668, 697, 758, 633, 520, 738, 691, 813, 428, 590, 706, 611, 655, 804, 663, 588, 877, 854, 654, 488, 836, 689, 717, 790, 629, 598, 736, 810, 841, 946, 795, 673, 776, 645, 646, 703, 608, 805, 746, 262, 730, 753, 878, 755, 837, 618, 594, 714, 543, 751, 263, 766, 817, 741, 534, 535, 536, 466  System Equipment Numbers: 74620447, 74620437, 74613625, 74613538, 74620341, 77839007, 73340583, 73326701, 77646555, 77559977, 77970948, 77427180, 72722066, 79132604, 72273457, 76100048, 77970800, 74620302, 72805859, 77228623, 80018072, 74611961, 77647014, 75486269, 78248928, 75486357, 76888124, 75581571, 78243622, 75423992, 77971015, 74817549, 76890243, 74352012, 77260210, 77213972, 78993356, 77347585, 75532686, 79152378, 79109118, 75533239, 74536213, 79152254, 77838972, 78150020, 78696158, 77559984, 76481840, 75848181, 79112403, 77905831, 80171865, 79042710, 77697096, 77213970, 74778339, 74778897, 73051857, 73717291, 78132484, 75305793, 72312081, 76890941, 76888973, 78365267, 76888978, 79464469, 72323525, 72323935, 74169336, 74620425, 76891009, 72417922, 79035318, 79035262, 78367631, 74584486, 74584491, 75615010, 75470091
510(k) numbers
["K181830"]
PMA numbers
not on file
Product code
JAA
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register