Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Perfusion Systems: Medtronic Vessel Cannulae: a) DLP 2 mm Vessel Cannula Blunt Tip, Model Number 30004; b) DLP 3 mm Vessel Cannula Blunt Tip, Model Number 30003

Recall number
Z-1799-2024
Recalling firm
Medtronic Perfusion Systems
Product
Medtronic Vessel Cannulae: a) DLP 2 mm Vessel Cannula Blunt Tip, Model Number 30004; b) DLP 3 mm Vessel Cannula Blunt Tip, Model Number 30003
Status
Open, Classified
Initiated
2024-03-19
Posted
2024-05-10
Terminated
not on file
Reason
Potential for unsealed sterile packing.
Root cause
Process control
Action
Medtronic issued an Urgent Medical Device Recall notice to its consignees on 03/19/2024 via letter (UPS). The notice explained the issue, potential risk, and requested the consignee identify and return the product. Consignee was directed to forward the notice to all affected parties. For questions, the consignee was directed to contact their Medtronic field representative.
Quantity
5719 units
Distribution
Worldwide distribution.
Lot, serial or model codes
a) Model Number 30004, GTIN 00613994646156, Lot Numbers: 2023041305; a) Model Number 30004, GTIN 00673978179841, Lot Numbers: 2023041300; a) Model Number 30004, GTIN 20613994646150, Lot Numbers: 2023041300, 2023041305, 202306C194; b) Model Number 30003, GTIN 20613994646174, Lot Numbers: 2023041298, 202306C221
510(k) numbers
["K810820"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1000116158
Firm city
Brooklyn Park
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register