Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical ASD Inc.: JELCO HYPODERMIC NEEDLE-PRO FIXED NEEDLE INSULIN SYRINGE: a) 1.0mL 28G X 1/2" (0.33mm x 13mm), REF 4428-1 b) 1.0mL 29G X 1/2" (0.33mm x 13mm), REF 4429-1

Recall number
Z-1798-2021
Recalling firm
Smiths Medical ASD Inc.
Product
JELCO HYPODERMIC NEEDLE-PRO FIXED NEEDLE INSULIN SYRINGE: a) 1.0mL 28G X 1/2" (0.33mm x 13mm), REF 4428-1 b) 1.0mL 29G X 1/2" (0.33mm x 13mm), REF 4429-1
Status
Terminated
Initiated
2021-05-17
Posted
2021-06-10
Terminated
2023-10-11
Reason
Specific models and lot numbers of the Jelco Hypodermic Needle-Pro Fixed Needle Insulin Syringe 29Gx1/2 odd number line graduation markings on the syringe barrel were skewed approximately 20 degrees upward.
Root cause
Process control
Action
The firm initiated the recall on May 17, 2021 by email. The firm is seeking the removal of the product down to the consumer level. For assistance ease contact Novasyte using the Email: smithsmedical-Jelco2021@novasyte.com The firm later realized they failed to include lot #40140906 for model #4429-1 and issued a second letter which was dated 5/27/2021 to customers involved in the first notification where this lot was excluded in the letter, as well as to new consignees who received the affected lot.
Quantity
1,432,200 units
Distribution
Worldwide distribution - US Nationwide and the countries of UK, Spain.
Lot, serial or model codes
a) REF 4428-1, Lot Numbers: 4046543 and 4062235 b) REF 4429-1, Lot Numbers: 4031846, 4031845, 4040734, 4043536, 4046545, 4046546, 4062239, 4062240, 4062238, 4062242, 4014096
510(k) numbers
["K063755"]
PMA numbers
not on file
Product code
MEG
Firm FEI
3012307300
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register