Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Sofamor Danek USA Inc: Medtronic INTREPID Spinal System SPACER, 32X23, 12 DEG, Size: 10MM, Part Number: REF 7961210; 12MM, REF 7961212; 14MM, REF 7961214; 16MM, REF 7961216; 18MM, REF 7961218; 20MM, REF 7961220; Rx only, STERILE R The INTREPID Spinal System is indicated for use with autogenous bone graft in patients with degenerative disc disease.

Recall number
Z-1797-2009
Recalling firm
Medtronic Sofamor Danek USA Inc
Product
Medtronic INTREPID Spinal System SPACER, 32X23, 12 DEG, Size: 10MM, Part Number: REF 7961210; 12MM, REF 7961212; 14MM, REF 7961214; 16MM, REF 7961216; 18MM, REF 7961218; 20MM, REF 7961220; Rx only, STERILE R The INTREPID Spinal System is indicated for use with autogenous bone graft in patients with degenerative disc disease.
Status
Terminated
Initiated
2009-02-06
Posted
2009-07-24
Terminated
2010-04-19
Reason
The firm received complaints regarding the use of the product in patients with poor bone quality, translational instability, difficulty implanting the cover plate and case of cover plate detachment.
Root cause
Device Design
Action
Medtronic Sofamor Danek USA, Inc. instructed Sales Representative on starting February 6, 2009 by phone to immediately return product from the field. The firm further issued a "Subject: Medtronic Intrepid Intervertebral Body Fusion Device Recall" notice dated April 2009. The recall notices were mailed directly to Risk Managers and Surgeons, with Recall Questionnaires and Surgeon Confirmation forms included respectively. For further questions, contact your Medtronic Sales Representative or the Medtronic Global Quality Department at 1-800-876-3133 extension 6333.
Quantity
506 units
Distribution
Worldwide Distribution -- United States, Australia, Germany, The Netherlands, and South Africa.
Lot, serial or model codes
All Lots.
510(k) numbers
["K080083"]
PMA numbers
not on file
Product code
MAX
Firm FEI
1000477302
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register