Philips North America Llc: Ingenia Ambition S - Model Number: 781359 Diagnostic MR system.

Recall number
Z-1796-2021
Recalling firm
Philips North America Llc
Product
Ingenia Ambition S - Model Number: 781359 Diagnostic MR system.
Status
Open, Classified
Initiated
2021-04-23
Posted
2021-06-04
Terminated
not on file
Reason
The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.
Root cause
Device Design
Action
On March 23, 2021, Philips issued an Urgent Medical Device Correction notice to customers via letter stating Philips identified an issue with specified MR systems. The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities A mechanical issue was detected on these hoisting interface blocks on top of the magnet. In the initial design of the hoisting blocks, the diameter to meet the required bolt-on interface surface for the swivel eye bolts, used as lifting hardware, did not meet the required specification. HOW TO IDENTIFY AFFECTED PRODUCTS Ingenia Ambition X (781356) and Ingenia Ambition S (781359) ACTION TO BE TAKEN BY CUSTOMER / USER No action is required by the customer, as the clinical use of the MR system is not affected. In case of a planned decommissioning or movement of the magnet, and this FCO is not installed yet, please contact your local Philips representative. ACTIONS PLANNED BY PHILIPS Via FCO781 00524 a mechanical solution will be rolled out to the affected MR systems. ***Updated November 2023*** Additional customers were identified and notified on August 22, 2023.
Quantity
156 MR systems
Distribution
Domestic: FL, GA, MI, MO, NH, NJ, NY, PA, SC, TX, VT. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Denmark, Finland, France, Germany, Iraq, Italy, Japan, Libya, Moldova, Netherlands, Oman, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom.
Lot, serial or model codes
REF 781359/ UDI-DI 00884838090057 Serial numbers: 48000 48001 48003 48004 48009 48005 48015 48017 48020 48011 48022 48012 48014 48028 48026 48037 48023 48043 48019 48160 48144 48180 48112 48052 ***Updated November 2023*** Added Serial Numbers 48051 48010 48038 48018 48039 48027 48122 48046 48036 48047 48133 48056 48169 48148 48154 48055 48128 48194 48173 48134 48040 48155 48186 48035 48181 48188 48198
510(k) numbers
["K181479"]
PMA numbers
not on file
Product code
LNH
Firm FEI
3016618143
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register