Skip to the content

FDA Medical Device Recalls (openFDA)

Merit Medical Systems, Inc.: PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT

Recall number
Z-1796-2020
Recalling firm
Merit Medical Systems, Inc.
Product
PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT
Status
Terminated
Initiated
2020-03-11
Posted
not on file
Terminated
2022-03-02
Reason
During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.
Root cause
Other
Action
On 03/11/2020, Urgent Product Recall Notices were mailed via second day service. The recalling firm requests that customers immediately stop using, quarantine, and return the affected lots. Further, they requested that customers make personnel within their organization aware of this field action. If the product was further distributed to other facilities, institutions, or manufacturers, customers were asked to ensure this notice was immediately shared with them and note the quantity distributed on the customer response form. Customers with additional questions were encouraged to call 801-208-4381.
Quantity
1932
Distribution
U.S.: NV, IL, VA, PA, CA, AR, MA, GA, OH, NJ, TX, FL, NC, SC, MI, NH, CO, AL, MN, KS, NM, AZ, ID, MD, NY, OR, DC, DE, MO, IN, WI, IA, TN. O.U.S.: Switzerland, Hong Kong, Spain, Portugal, Canada, Singapore, United Arab Emirates, Mexico, United Kingdom, France, Japan, Germany, Italy
Lot, serial or model codes
SDRB-REG-LT: Lot(UDI): H1559517(10884450363673) and H1727668(00884450363676). SDRB-REG-RT: Lot(UDI): H1666548(10884450363659) and H1723973(10884450363659)
510(k) numbers
["K180723"]
PMA numbers
not on file
Product code
DXC
Firm FEI
1721504
Firm city
South Jordan
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register