Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp.: Integra Mozaik Putty 2.5cc, 5 cc, 10 cc Product Usage: The product is a bone graft substitute composed of collagen (Type 1) and ceramic (IP3-Tricalcium phosphate) matrix in putty and strip forms. The product is intended for use as a bone void filler to fill voids or gaps of the skeletal system in the extremities, spine. and pelvis not intrinsic to the stability of the bony structure. Integra Mozaik is also indicated for use in the treatment of surgically treated osseous defects or osseous defects created from traumatic injury to the bone. Following placement in the bone void or a defect Integra Mozaik is reabsorbed and replaced with bone during the healing process.

Recall number
Z-1796-2012
Recalling firm
Integra LifeSciences Corp.
Product
Integra Mozaik Putty 2.5cc, 5 cc, 10 cc Product Usage: The product is a bone graft substitute composed of collagen (Type 1) and ceramic (IP3-Tricalcium phosphate) matrix in putty and strip forms. The product is intended for use as a bone void filler to fill voids or gaps of the skeletal system in the extremities, spine. and pelvis not intrinsic to the stability of the bony structure. Integra Mozaik is also indicated for use in the treatment of surgically treated osseous defects or osseous defects created from traumatic injury to the bone. Following placement in the bone void or a defect Integra Mozaik is reabsorbed and replaced with bone during the healing process.
Status
Terminated
Initiated
2011-10-11
Posted
2012-06-12
Terminated
2013-02-14
Reason
The Instructions for Use {HIFU} {Putty Part Number 7700713001 and Strip Part Number 770151300 l for CE marked Mozaik Strip and Putty products contains an incorrect translation of the term "bone marrow aspirate". The term is incorrectly translated in Spanish and French,
Root cause
Employee error
Action
Integra LifeSciences sent a Field Safety Notice letter dated October 11, 2011 to their sole foreign consignee. The letter identified the affected products, problem and actions to be taken. The consignee was asked to remove the existing Instructions for Use (IFU) and replace it with the corrected IFU. The consignee was instructed to complete and return the attached Distributor Rework and Acknowledgment Forms provided with the customer notification letter as documentation that instructions were followed and all required actions were completed. For any questions, call: +33 (0)4 37 47 5900 , fax: +33 (0)4 37 47 5999; or email: carole.robin@integralife.com.
Quantity
42 units
Distribution
Worldwide Distribution
Lot, serial or model codes
Putty Part Number 7700713001 510 K- K062353 Device Listing Number - E604163  Integra MOZAIKTM Putty 2.5cc Reference No. - PTY10125ITL Batch number - 105BR02l5949  Integra MOZAIKTM Putty 5cc Refernce No. - PTY10155ITL Batch No. - 105BR0216024  Integra MOZAIKTM Putty lOcc Reference No.- PTY10210ITL Batch No. - 105BR0215956
510(k) numbers
["K062353"]
PMA numbers
not on file
Product code
MQV
Firm FEI
3003418325
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register