bioMerieux, Inc.: PREVI Isola Insrument and PREVI Isola Printer spares PSUs (power supply units), P1028888-001 PSU (used in Zebra LP2824 Plus printers) PSUs with the following date code; 0639XX through 1253XX (YYWWXX, where YY=Year, and WW=week).

Recall number
Z-1793-2020
Recalling firm
bioMerieux, Inc.
Product
PREVI Isola Insrument and PREVI Isola Printer spares PSUs (power supply units), P1028888-001 PSU (used in Zebra LP2824 Plus printers) PSUs with the following date code; 0639XX through 1253XX (YYWWXX, where YY=Year, and WW=week).
Status
Terminated
Initiated
2020-01-24
Posted
not on file
Terminated
2023-01-11
Reason
The recall has been initiated due to potential for overheating or a fire hazard associated with the Zebra LP2824 Plus printer PSU (power supply unit) located within the PREVI Isola instrument.
Root cause
Nonconforming Material/Component
Action
The firm initiated the recall by letter on 01/15/2020. The Urgent Product Correction Notice instructed the customer the to provide the letter to all appropriate personnel within their organization. The letter instructs the personnel to check their inventory against the serial numbers listed in Appendix A of the letter. The letter also details that a bioMerieux Field Service Engineer will be contacting the customer to schedule a service appointment and if necessary, replace the PSU.
Quantity
not on file
Distribution
Worldwide distribution. US nationwide including AU, BE, CH, DE, ES, FR, IT, HK/CN; MX, NL, RU, SE/FI, ZA.
Lot, serial or model codes
Instrument Serial Number  AS180-00028 1035196  AS180-00067 1007580  AS180-00084 2055833  AS180-00105 2025461  AS180-00138 2025461  AS180-00179 1039938  AS180-00217 1033454  AS180-00248 1029262  AS180-00278 1031886  AS180-00300 1033400  AS180-00306 1034034  AS180-00326 1034190  AS180-00336 1041241  AS180-00339 2054793  AS180-00357 2055775  AS180-00381 1073849  AS180-00385 1054234  AS180-00398 1052053  AS180-00402 2055620  AS180-00405 1063739  AS180-00418 1042188  AS180-00420 1040343  AS180-00428 1030890  AS180-00453 1031472  AS180-00468 1043640  AS180-00476 1029272  AS180-00478 1053432  AS180-00479 1011691  AS180-00480 5001554  AS180-00481 1039588  AS180-00482 1034363  AS180-00483 1029990  AS180-00484 1049298  AS180-00485 AU01  AS180-00486 1047965  AS180-00487 1074074  AS180-00488 1000283  AS180-00489 1035594  AS180-00490 1047965  AS180-00491 1035549  AS180-00492 1070878  AS180-00493 1090075  AS180-00494 1039290  AS180-00495 1029148  AS180-00500 1007884  AS180-00509 1065153
510(k) numbers
not on file
PMA numbers
not on file
Product code
not on file
Firm FEI
3002769706
Firm city
Durham
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register