Skip to the content

FDA Medical Device Recalls (openFDA)

Argon Medical Devices, Inc: 0.035"(.89mm)x150 cm guidewire. Model 114135150. Soft Tip 3.5cm. REF/UDI: 114135150/(01)20886333008074. RxOnly.STERILE EO

Recall number
Z-1792-2020
Recalling firm
Argon Medical Devices, Inc
Product
0.035"(.89mm)x150 cm guidewire. Model 114135150. Soft Tip 3.5cm. REF/UDI: 114135150/(01)20886333008074. RxOnly.STERILE EO
Status
Terminated
Initiated
2019-12-04
Posted
2020-05-07
Terminated
2024-04-01
Reason
The stiffness of the soft end of the guidewire caused tissue perforation.
Root cause
Employee error
Action
On 12/04/2019, the firm sent Medical Device Recall letter by email to their customer/ distributors. Customers are instructed to inspect and quarantine and return any affected product remaining at their facility utilizing the instruction on the Recall Response Form. Customers are instructed to complete the response acknowledgement form included with the letter and return by fax, mail or email as soon as possible. Customers were also instructed to pass the notice on to all those who need to be aware within their organization or to any organization to which the potentially affected devices have been transferred.
Quantity
1190 units
Distribution
No US consignees. OUS - Sweden, Norway, Denmark, Greece, Slovenia, Turkey, Hong Kong, England
Lot, serial or model codes
Guidewire Model 114135150; LOT numbers: 11248718, 11251204, 11260495. 11268741, 11242667, 11241250 and 11239540
510(k) numbers
["K160785"]
PMA numbers
not on file
Product code
DQX
Firm FEI
1625425
Firm city
Athens
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register