FDA Medical Device Recalls (openFDA)
TriMed Inc.: TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm, 65mm, TI, Ref: IMN3.6-65 T, Non Sterile, MD, Rx Only
- Recall number
- Z-1791-2024
- Recalling firm
- TriMed Inc.
- Product
- TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm, 65mm, TI, Ref: IMN3.6-65 T, Non Sterile, MD, Rx Only
- Status
- Open, Classified
- Initiated
- 2024-03-19
- Posted
- 2024-05-10
- Terminated
- not on file
- Reason
- Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.
- Root cause
- Process control
- Action
- On 4/3/24, TriMed Inc., issued a "Urgent: Medical Device Recall", to affected consignees via E-Mail. TriMed ask consignees to take the following actions: " If you have the affected product, discontinue further use and quarantine products prior to return. " If a successful implantation was completed no further action is necessary. Only standard monitoring of the patient is required after successful implantation. " Please share this notice with all personnel that may utilize the affected products. Please share this notification to any locations where the affected product may have been transferred to. " Conduct a physical count of the affected product in your possession and record the count on attached response form. Once the response form is completed, email to: Quality@trimedortho.com. Please respond even if you have no affected products in your inventory. " Once the response form is completed and emailed, you will receive an RGA (Return Good Authorization form) as well as a link to generate a label to return the devices via email. Please package the devices, print the return label and ship via FedEx.
- Quantity
- 142 units
- Distribution
- US Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IL, LA, MA, MI, NC, NJ, NV, OH, OK, PA, TN, TX, UT, VA, WA, WI.
- Lot, serial or model codes
- UDI: 00842188127878/ Lot # QM23086
- 510(k) numbers
- ["K234040"]
- PMA numbers
- not on file
- Product code
- HWC
- Firm FEI
- 3001236812
- Firm city
- Santa Clarita
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28