Childrens Medical Ventures: Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.
- Recall number
- Z-1789-2014
- Recalling firm
- Childrens Medical Ventures
- Product
- Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.
- Status
- Terminated
- Initiated
- 2014-04-29
- Posted
- 2014-06-13
- Terminated
- 2015-04-06
- Reason
- The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.
- Root cause
- Component design/selection
- Action
- Philips sent an Urgent Medical Device Correction letter dated April 29, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. This Customer Information is intended to inform you about: 1.what the problem is and under what circumstances it can occur 2.the actions that you as a customer can take to minimize the effect of the problem 3.the actions planned by Philips to correct the problem. If you need any further information or support concerning this issue, please contact your local Philips representative: For further questions please call (412) 380-8881. Philips apologizes for any inconveniences caused by this problem.
- Quantity
- 112
- Distribution
- Worldwide Distribution - USA including the states of MI, WI, SC, FL, TX, KS, MA, AZ, UT, WV, IN, MN, NV and VA., and the countries of Australia, Belgium and Canada.
- Lot, serial or model codes
- 989805603611 / 1014557, 989805603631 / 1014559, 989805604511 / 1028969, 989805604521 / 1028970, 989805604531 / 1028971, 989805604631 / 1030271, 989805605051 / 1038140, 989805605061 / 1038141, 989805608801 / U1014557, 989805608811 / U1014559, 989805608831 / U1028969, 989805608841 / U1028970, 989805608851 / U1038140, 989805604231 / 1023384
- 510(k) numbers
- ["K061256"]
- PMA numbers
- not on file
- Product code
- NPF
- Firm FEI
- 3007056120
- Firm city
- Monroeville
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28