Skip to the content

FDA Medical Device Recalls (openFDA)

General Electric Med Systems LLC: GE Healthcare Definium 8000 Digital Radiographic Systems : GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model Number: 5131070. The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic images of human anatomy. It is not intended for mammographic use.

Recall number
Z-1789-2012
Recalling firm
General Electric Med Systems LLC
Product
GE Healthcare Definium 8000 Digital Radiographic Systems : GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model Number: 5131070. The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic images of human anatomy. It is not intended for mammographic use.
Status
Terminated
Initiated
2006-05-31
Posted
2012-06-08
Terminated
2012-07-31
Reason
An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath
Root cause
Device Design
Action
GE Healthcare sent an "URGENT SAFETY INFORMATION" letter dated May 31, 2006 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included new operator instructions and warning labels.
Quantity
3,796 units
Distribution
Worldwide Distribution.
Lot, serial or model codes
All serial numbers
510(k) numbers
["K012389"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register