FDA Medical Device Recalls (openFDA)
General Electric Med Systems LLC: GE Healthcare Definium 8000 Digital Radiographic Systems : GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model Number: 5131070. The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic images of human anatomy. It is not intended for mammographic use.
- Recall number
- Z-1789-2012
- Recalling firm
- General Electric Med Systems LLC
- Product
- GE Healthcare Definium 8000 Digital Radiographic Systems : GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model Number: 5131070. The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic images of human anatomy. It is not intended for mammographic use.
- Status
- Terminated
- Initiated
- 2006-05-31
- Posted
- 2012-06-08
- Terminated
- 2012-07-31
- Reason
- An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath
- Root cause
- Device Design
- Action
- GE Healthcare sent an "URGENT SAFETY INFORMATION" letter dated May 31, 2006 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included new operator instructions and warning labels.
- Quantity
- 3,796 units
- Distribution
- Worldwide Distribution.
- Lot, serial or model codes
- All serial numbers
- 510(k) numbers
- ["K012389"]
- PMA numbers
- not on file
- Product code
- MQB
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28