Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation: STERIS Fine Traction Device, one unit per box Product Usage: The Fine Traction Device is used to provide traction and enable fine adjustments to traction and orientation of the leg during orthopedic procedures.

Recall number
Z-1786-2013
Recalling firm
Steris Corporation
Product
STERIS Fine Traction Device, one unit per box Product Usage: The Fine Traction Device is used to provide traction and enable fine adjustments to traction and orientation of the leg during orthopedic procedures.
Status
Terminated
Initiated
2013-05-20
Posted
2013-07-22
Terminated
2013-08-28
Reason
A bearing used in the assembly of the Fine Traction Device did not perform successfully in pre-release reliability testing; a replacement bushing passed reliability testing and was substituted per a change order. The change order authorized use of the substitute bushing in assembly of the Fine Traction Accessory; however, inadvertently the new part was not implemented. During complaint evaluati
Root cause
Component change control
Action
STERIS Account Managers notified affected customers via onsite visits. All consignees were notified as of 5/29/2013. Customers were instructed to return affected product for a replacement unit.
Quantity
16 units
Distribution
USA Nationwide Distribution in the states of CO, KY, LA, VA, and WI. The affected product was also distributed to Canada.
Lot, serial or model codes
The following serial/lot numbers are affected by this recall:  0404913150, 0404913151, 0404913152, 0404913153, 0408713001, 0408713002, 0408713003, 0408713004, 0408713005, 0408713006, 0409813097, 0409813098, 0409813099, 0409813100, 0410113001, 0410113002, 0410713024, 0410713025, 0410713026, 0410713027, 0410713028, and 0410713029.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HST
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register