Skip to the content

FDA Medical Device Recalls (openFDA)

Captiva Spine, Inc: Accessories for the Captiva Spine FuseLOX Lumbar Cage. The Trial product is a reuseable component of the Captiva Spine FuseLOX Lumbar product. It is not generally sold as an individually packaged product. Trial products are shipped as a component of the FuseLOX Lumbar Caddy which is placed in the FuseLox Lumbar Sterilization tray. The sterilization tray is distributed on consignment. FPT - F is FuseLOX, P is Plif, and T is Trial. Products subject to recall: FuseLOX Lumbar Trials Trial - 7mm x 25mm Part Number: FPT2507 Lot Number 8110001. Trial - 8mm x 25mm Part Number FPT2508 Lot Number 8110002. Trial - 9mm x 25mm Part Number: FPT2509 Lot Number 8110003. Trial - 10mm x 25mm Part Number: FPT2510 Lot Number 8110004. Trial - 11mm x 25mm Part Number: FPT2511 Lot Number 8110005. Trial - 12mm x 25mm Part Number: FPT2512 Lot Number 8110006. Trial - 13mm x 25mm Part Number FPT2513 Lot Number 8110007. Trial - 14mm x 25mm Part Number FPT2514 Lot Number 8110008. The FuseLOX Lumbar Cage implant is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The FuseLOX Lumbar Cage Implant is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine.

Recall number
Z-1785-2012
Recalling firm
Captiva Spine, Inc
Product
Accessories for the Captiva Spine FuseLOX Lumbar Cage. The Trial product is a reuseable component of the Captiva Spine FuseLOX Lumbar product. It is not generally sold as an individually packaged product. Trial products are shipped as a component of the FuseLOX Lumbar Caddy which is placed in the FuseLox Lumbar Sterilization tray. The sterilization tray is distributed on consignment. FPT - F is FuseLOX, P is Plif, and T is Trial. Products subject to recall: FuseLOX Lumbar Trials Trial - 7mm x 25mm Part Number: FPT2507 Lot Number 8110001. Trial - 8mm x 25mm Part Number FPT2508 Lot Number 8110002. Trial - 9mm x 25mm Part Number: FPT2509 Lot Number 8110003. Trial - 10mm x 25mm Part Number: FPT2510 Lot Number 8110004. Trial - 11mm x 25mm Part Number: FPT2511 Lot Number 8110005. Trial - 12mm x 25mm Part Number: FPT2512 Lot Number 8110006. Trial - 13mm x 25mm Part Number FPT2513 Lot Number 8110007. Trial - 14mm x 25mm Part Number FPT2514 Lot Number 8110008. The FuseLOX Lumbar Cage implant is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The FuseLOX Lumbar Cage Implant is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine.
Status
Terminated
Initiated
2012-04-27
Posted
2012-06-07
Terminated
2013-12-24
Reason
Captiva Spine, Inc. initiated a recall of FuseLOX Lumbar Implants, Trials, and Inserters after FDA advised the firm that there is inadequate support for implemented design changes to the cleared device Pivotec Lumbar Interbody Fusion Device.
Root cause
Finished device change control
Action
Captiva Spine sent a Voluntary Correction Notification letter dated May 4, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to locate and return affected devices. Customers were also instructed to complete and return a Tracking/Verification Form. For questions customers were instructed to call 877-772-5571. For questions regarding this recall call 561-277-9480.
Quantity
54 units
Distribution
USA (nationwide) including FL,TX, UT, and NJ.
Lot, serial or model codes
Lot Number: 8110001 thru 8110008.
510(k) numbers
not on file
PMA numbers
not on file
Product code
MAX
Firm FEI
3006082533
Firm city
Jupiter
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register