Beckman Coulter Inc.: DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065. The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
- Recall number
- Z-1784-2025
- Recalling firm
- Beckman Coulter Inc.
- Product
- DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065. The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
- Status
- Open, Classified
- Initiated
- 2025-03-20
- Posted
- 2025-05-15
- Terminated
- not on file
- Reason
- Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.
- Root cause
- Software design
- Action
- On or about 03/20/2025, the firm sent via email or postal mail an Urgent Medical Device Recall letters informing customers that Beckman Coulter has determined that the DxC 500 AU Clinical Chemistry Analyzer may not run a requested calibration order for a two-part reagent and subsequently prevents any new calibration order from being placed. For two-part reagents, when reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. The analyzer will not allow additional calibration orders to be requested for any assays. Customers are instructed to: . Do not place calibration orders when the analyzer is processing samples. . Refer to the DxC 500 AU IFU Reagent Overview section for instructions on Monitoring the Status of Reagents. . Beckman Coulter recommends sharing the content of this letter with their laboratory and/or Medical Director. . If they experience this issue on software version 1.3, 1.4, 1.4.1 or 1.4.2, contact Beckman Coulter for service. . Beckman Coulter recommends posting this Recall Letter on or near the affected systems until the corrections have been made. Resolution: . This has been resolved in software version 1.4.3 in December of 2024. . Your Beckman Coulter service representative will contact you to schedule the software upgrade. For questions - contact Customer Support Center at: . From our website: http://www.beckmancoulter.com . Troubleshooting Hotline: (800) 854-3633
- Quantity
- not on file
- Distribution
- U.S: AL, CA, GA, ID, IN, LA, MA, MI, MN, MT, NC, NV, PA, PR, TN , TX, and WI O.U.S.: Australia, Belgium, Bosnia and Herzegovina, Canada, Chile, Croatia, Egypt, Ethiopia, France, Germany, Ghana, India, Italy, Korea, Lebanon, Libya, Malaysia, Netherlands, North Macedonia, Oman, Paraguay, Philippines, Poland, Puerto Rico, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, and Tunisia
- Lot, serial or model codes
- Catalog Number: C63519 UDI-DI code: 14987666545058 Serial Numbers: Catalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers:
- 510(k) numbers
- ["K220977"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28