NRT X-RAY A/S: Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
- Recall number
- Z-1784-2024
- Recalling firm
- NRT X-RAY A/S
- Product
- Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
- Status
- Open, Classified
- Initiated
- 2024-04-17
- Posted
- 2024-05-07
- Terminated
- not on file
- Reason
- X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
- Root cause
- Device Design
- Action
- On 04/17/24, correction notices were emailed to consignees who were asked to do the following: 1) If you experience the described uncontrolled motion, stop using the device and contact your service provider to have your device repaired. 2) The correction notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. 3) Complete and return the response form to either support@nrtxray.com or servicesupport@alpha-imaging.com A firmware update will be made to all motor controllers in the system. The update will be performed by the Service Provider. Customers with recall-related questions can call 1-800-331-7327.
- Quantity
- 35
- Distribution
- US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI.
- Lot, serial or model codes
- REF/UDI-DI/Serial Numbers: 03400000/05713464000527/627002, 627004, 627007, 639001, 639002, 639003, 639004, 639007, 639010, 647001, 647004, 647005, 648001, 651007, 658002, 658003, 658006, 658010, 666008, 675001, 675004, 682006, 682010, 682011, 696003, 696004, 696006, 696007, 696012. 03400010/05713464000121/647009, 651002, 651006, 666009, 682003, 682023
- 510(k) numbers
- ["K173631"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 1000188474
- Firm city
- Hasselager
- Firm state
- not on file
- Firm country
- Denmark
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28