FDA Medical Device Recalls (openFDA)
Instrumentation Laboratory Co.: HemosIL RecombiPlasTin 2G, Part Number 0020002950 (8 ml size). The product is used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Anticoagulant Therapy.
- Recall number
- Z-1784-2013
- Recalling firm
- Instrumentation Laboratory Co.
- Product
- HemosIL RecombiPlasTin 2G, Part Number 0020002950 (8 ml size). The product is used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Anticoagulant Therapy.
- Status
- Terminated
- Initiated
- 2013-04-09
- Posted
- 2013-07-22
- Terminated
- 2015-05-06
- Reason
- Instrumentation Laboratory is recalling HemosIL RecombiPlasTin 2G (RPT2G), Part No. 0020009580 (8 mL Size) due to being Out of Specification.
- Root cause
- Nonconforming Material/Component
- Action
- Between April 9, 2013 and April 11, 2013, Instrumentation Laboratory distributed "Important Product Correction" notification letters to their U.S. and foreign customers informing them that the on-board instrument stability claim has been reduced to 3 days for all current and future product lots of HemosIL RecombiPlasTin 2G. A second notification to non-respondents was initiated and completed on May 7, 2013. Technical questions were referred to 1-800-678-0710, Option # 2.
- Quantity
- Kits per product:1)11080, 2)11603, 3)11487, 4)11483, 5)11290, 6)11562, 7)11535, 8)11577, 9)11505, 10)11708,11)11579,12)11580, 13&14) not distributed
- Distribution
- Worldwide Distribution- US (nationwide) including states of: MA, ME, CT, NJ, NY, PA, MD, VA, NC, SC, GA, FL, AL, TN, MS, KY, OH, IN, MI, IA, WI, MN, SD, MT, IL, MO, KS, NE, LA, AR, OK, TX, CO, ID, AZ, CA, OR, WA, AK, RI, VT and the countries of: Angola, Argentina, Australia, Austria, Belgium, Bermuda, Brazil, Bulgaria, Canada, Caribbean, Chile, Colombia, Costa Rica, Czech, Denmark, Dubai, Ecuador, Egypt, France, Germany, Gibraltar, Great Britain, Greece, Guatemala, Hungary, India, Indonesia, Iran, Israel, Italy, Jamaica, Jordan, Korea, Kosovo, Kuwait, Lebanon, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, Norway, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad, Turkey, Ukraine, Uruguay, Vietnam, and Japan.
- Lot, serial or model codes
- Part Number applies to all: 0020002950 - all product lots. Lot Numbers: 1) N0716706 Exp. Date 5/31/2013, 2) N0817321 Exp. Date 5/31/2013, 3) E0917396 Exp. Date 9/30/2013, 4)) N0917378 Exp. Date 9/30/2013, 5) N1118029 Exp. Date 10/31/2013, 6) N0128669 Exp. Date 1/31/2014, 7) N0329291 Exp. Date 2/28/2014, 8) N0429596 Exp. Date 3/31/2014, 9) N0520056 Exp. Date 5/31/2014,10) E0720822 Exp. Date 7/31/2014, 11) N0821174 Exp. Date 8/31/2014, 12) N091388 Exp. Date 9/30/2014, 13) N0921522 Exp. Date 9/30/2014, 14) N0132676 Exp. Date 1/31/2015 and 15) N1222567 Exp. Date 1/31/2015.
- 510(k) numbers
- ["K070005"]
- PMA numbers
- not on file
- Product code
- GJS
- Firm FEI
- 1217183
- Firm city
- Bedford
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28