FDA Medical Device Recalls (openFDA)
Philips Respironics, Inc.: Wisp and Wisp Youth Nasal Mask
- Recall number
- Z-1783-2022
- Recalling firm
- Philips Respironics, Inc.
- Product
- Wisp and Wisp Youth Nasal Mask
- Status
- Open, Classified
- Initiated
- 2022-09-06
- Posted
- 2022-09-30
- Terminated
- not on file
- Reason
- No contraindication/warning regarding magnetic components in CPAP masks for patients with implantable devices or metallic splinters in the eyes.
- Root cause
- Error in labeling
- Action
- The firm started notifying customers of this recall event via telephone on 9/6/2022. Additionally, customers were mailed an URGENT Medical Device Correction Letter dated 9/9/2022 that communicates that the firm is updating the Contraindications and Warning language for the Amara View Full Face Mask, DreamWisp Nasal Mask, DreamWear Full Face Mask, Wisp and Wisp Youth Nasal Mask, and Therapy Mask 3100 NC/SP due to the presence of magnets that may affect implanted medical devices and metallic splinters in the eye. All units of these masks are subject to this recall action. If a patient, a household member, caregiver, or bedpartner that is in close vicinity to patients using the mask is equipped with one of the medical devices or metallic splinters indicated in the updated Contraindications and Warning language, patients are to stop use of the affected mask and properly dispose of it once an alternative mask without magnets has been obtained after consultation with their healthcare provider. If the patient or others around them does not possess an implanted device or metal splinters included in the updated contraindications language, no action is needed. Users and those in the vicinity of the patient while using the mask must ensure the mask is kept at least six inches from medical implants and metallic splinters in the eye. Customers with any questions are to contact Philips Customer Care Solutions at 1-800-345-6443. The firm issued a press release on 9/6/22 and can be found at the following link: https://www.philips.com/a-w/about/news/archive/standard/news/press/2022/20220906-philips-respironics-alerts-customers-worldwide-of-updated-instructions-and-labeling-of-specific-sleep-therapy-masks-that-contain-magnetic-headgear-clips-due-to-potential-risk-of-serious-injury.html. A second Urgent Medical Device Correction Letter was sent to customers on 12/1/2022. Should DMEs, distributors, and prescribers decide to continue distributing masks and mask accessories with magnet
- Quantity
- 6,237,558 units
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming, & District of Columbia. The countries of Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bermuda, Bolivia, Bosnia-Herz., Brazil, Brunei Darussal, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Guadeloupe, Guam, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Korea, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mexico, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, St. Pier, Miquel., Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, United States, Uruguay, Utd. Arab. Emir., Uzbekistan, & Zimbabwe.
- Lot, serial or model codes
- All Units; Model No.: 1094050, 1094051, 1094053, 1094060, 1094061, 1094062, 1094070, 1094071, 1094072, 1094073, 1109298, 1113954, 1117119, 1118060, 1118061, 1118062, 1118063, 1118064, 1118065, 1118066, 1118067, 1120434, 1137360, 1137361, 1137366, 1137367, 1137368, 1137369, 1137370, 1138989, 1138991, 1138992, 1138993, 1138994, 1138995, 1138996, 1139009, 1148366, 1094051RTL.
- 510(k) numbers
- ["K121631","K140268"]
- PMA numbers
- not on file
- Product code
- BZD
- Firm FEI
- 2518422
- Firm city
- Murrysville
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28