Skip to the content

FDA Medical Device Recalls (openFDA)

TP Orthodontics, Inc.: Jasper Vektor, Class II Correction Appliance, Part No. 610-500 used for correcting class 2 malocclusions by applying light forces on the teeth using a tube, arch wire and stops.

Recall number
Z-1783-2018
Recalling firm
TP Orthodontics, Inc.
Product
Jasper Vektor, Class II Correction Appliance, Part No. 610-500 used for correcting class 2 malocclusions by applying light forces on the teeth using a tube, arch wire and stops.
Status
Terminated
Initiated
2018-05-01
Posted
not on file
Terminated
2020-04-15
Reason
The firm has received 106 complaints of the device breaking in the patient's mouth. Device breakage may cause injury to the patient.
Root cause
Device Design
Action
On May 3, 2018, the firm sent an Urgent Medical Device Recall letter to customers. The letter advised customers that the firm was voluntarily recalling Generation 1 Jasper Vektor Class 2 Correction Appliances due to receiving 106 customer complaints that the device had broken in the patient's mouth during placement and usage. Customers were instructed to do the following: 1. Locate any of the affected lots identified in the table below and quarantine the product. 2. Return the Generation 1 Jasper Vektor for a replacement Jasper Vektor Generation 2 or refund, following the instructions below: Contact TP Orthodontics, Inc. at 800-338-8856 or info@tportho.com for a Return Goods Authorization and pre-paid shipping label. Alternately, use the carrier of your choice with available tracking and send to: TP Orthodontics, Inc. ATTN: Recall Coordinator 100 Center Plaza La Porte, IN 46350 3. Email, mail, or fax the attached response letter to TP Orthodontics, Inc. In addition to sending the letter back, you may call us with any questions or concerns. EMAIL: info@tportho.com FAX: 219-324-3029 MAIL: TP Orthodontics, Inc. ATTN: Recall Coordinator 100 Center Plaza La Porte, IN 46350 4. If you have any further questions or concerns after reviewing this notice, please contact TPO Customer Care Team 800-388-8856 or info@tportho.com
Quantity
108
Distribution
Distributed to accounts in 20 states: AL, CA, CO, FL, HI, ID, IL, MA, ME, MI, NJ, NV, NY, OH, OR, PA, SC, TX, UT, and VA. Foreign distribution to United Arab Emirates, Australia, Belgium, Brazil, Canada, Germany, Spain, Finland, France, Great Britain, Hungary, Ireland, Italy, Japan, Mexico, Netherlands, New Zealand, Portugal, Singapore, and South Africa.
Lot, serial or model codes
15915000, 17615000, 17715000, 18015000, 18115000, 18915000, 20315000, 23915000, 24515000, 25715000, 25815000, 26515000, 29415000, 29315000, 32915000
510(k) numbers
not on file
PMA numbers
not on file
Product code
ECO
Firm FEI
1815540
Firm city
La Porte
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register