Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Boston Scientific Imager II Angiographic Catheter, M001316581 (5-pack outer package UPN), M001316580, single unit, inner package UPN)

Recall number
Z-1781-2017
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific Imager II Angiographic Catheter, M001316581 (5-pack outer package UPN), M001316580, single unit, inner package UPN)
Status
Terminated
Initiated
2016-12-13
Posted
2017-03-23
Terminated
2017-11-22
Reason
One lot of Imager II Angiographic Catheters is being recalled due to packaging mislabeling. The product may be labeled as Imager II Contra 2 curve catheters, when in fact the package contains Contra curve catheters.
Root cause
Employee error
Action
Customers were sent an Urgent Medical Device Removal letter, dated 12/5/2016, via overnight mail delivery beginning 12/15/2016. The letter identified affected product, described the issue, and asked for product to be returned. Removal instructions were provided and a Reply Verification Tracking Form was asked to be returned. Questions can be directed to Nicole Pshon, Field Action Manager at 763-494-1133 or BSCFieldActionCenter@bsci.com
Quantity
26 boxes of 5 (130 singles)
Distribution
AZ, FL, MO, NJ, NY, OH, PA, and TX.
Lot, serial or model codes
Lot: 00116127, Expiration Date: September 30, 2018
510(k) numbers
["K121694"]
PMA numbers
not on file
Product code
DQO
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register