National Biological Corp: Dermalight 80 Model Number UVB-080, one unit per box. The Dermalight 80 is a Phototherapy light unit that provides ultraviolet radiation therapy for diagnosed skin disorders such as Psoriasis and Vitiligo.

Recall number
Z-1778-2013
Recalling firm
National Biological Corp
Product
Dermalight 80 Model Number UVB-080, one unit per box. The Dermalight 80 is a Phototherapy light unit that provides ultraviolet radiation therapy for diagnosed skin disorders such as Psoriasis and Vitiligo.
Status
Terminated
Initiated
2013-04-09
Posted
2013-07-19
Terminated
2014-02-24
Reason
The incorrect manual may have been packaged with this product. UVB-080 devices may have included the manual for a UVA-083 device.
Root cause
Labeling mix-ups
Action
The firm, National Biological, sent an "VOLUNTARY PRODUCT RECALL NOTIFICATION" letter dated May 10, 2013 to its consignees/customers by UPS or other delivery service with a signature required. The letter described the product, problem and actions to be taken. Consignees/customers who do not respond to the letter will be contacted via telephone. The customers were instructed to check within your household to determine if you have any of the listed product; immediately destroy the incorrect User Manual, it is not necessary to return the product; record the Customer Name and or the Dermalight-80 serial number and complete and return the enclosed Verification Form, even if you do not have any User Manuals to destroy. Please contact your National Biological Corporation sales representative if you have any questions regarding this recall, any of our products, or would like assistance with the recall call 216-831-0600 or 800-338-5045 email: infor@PhototherapyExperts.com.
Quantity
124 units
Distribution
Worldwide distribution: US (nationwide) including states of: AL, CA, CO, CT, FL, GA, IA, ID, IL, LA, MA, MN, MO, MS, NE, NJ, NY, OH, PA, SC, TN, TX, VA, WA, and WI; and countries of: Canada, Brazil, Malaysia, and Mexico.
Lot, serial or model codes
The following batch numbers and serial numbers are affected by this recall: Batches 7839 and 7951, Serial numbers D80-12748 to D80-12871.
510(k) numbers
["K903660"]
PMA numbers
not on file
Product code
FTC
Firm FEI
1521608
Firm city
Beachwood
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register