Collagen Matrix Inc: Collagen Dental Wound Dressing - Bovine Dermis Tap, 25mm x 75mm Product was distributed under the following brand names: 1) Cytoplast RTMTAPE, 25mm x 75mm 2) GIBSON Healthcare, 25mm x 75mm 3) MaxxTape, 2.5cm x 7.5cm x 0.1cm 4) NEOTAPE, 25mm x 75mm 5) Karl Schumacher TAPE, 2.5cm x 7.5cm x 1mm

Recall number
Z-1777-2022
Recalling firm
Collagen Matrix Inc
Product
Collagen Dental Wound Dressing - Bovine Dermis Tap, 25mm x 75mm Product was distributed under the following brand names: 1) Cytoplast RTMTAPE, 25mm x 75mm 2) GIBSON Healthcare, 25mm x 75mm 3) MaxxTape, 2.5cm x 7.5cm x 0.1cm 4) NEOTAPE, 25mm x 75mm 5) Karl Schumacher TAPE, 2.5cm x 7.5cm x 1mm
Status
Open, Classified
Initiated
2022-08-03
Posted
2022-09-23
Terminated
not on file
Reason
The packaging may not be sealed, which could compromise sterility.
Root cause
Packaging change control
Action
On August 3, 2022, Collagen Matrix began informing the distributors (initial consignees) to place the lot in question on ship hold. On August 8, 2022, Collagen Matrix began informing distributors by email to return the concerned products. Consignees are asked to return any recalled product to the recalling firm and complete the Certificate of Acknowledgement within 15 days of receipt and return it via email.
Quantity
1336 units
Distribution
US distribution to distributors in IL, IN, MI, NC, OH, and TX
Lot, serial or model codes
1) UDI (1 unit package): +D765RTMTAPE100/$$3250131E1C0070AJ; (10 units package): +D765RTMTAPE101/$$3250131E1C0080AK Cytoplast RTMTAPE, Lot E1C0070A 2) UDI (1 unit package): (01)00366975012616(17)250131(10)BDDWTU211002; (10 units package): (01)10366975012613(17)250131(10)BDDWTU211002 GIBSON Healthcare, REF 5251-194; Lot BDDWTU211002 3) UDI (1 unit package): (01)00813954023571(17)250131(10)BDDWTU211002; (10 units package): (01)10813954023578(17)250131(10)BDDWTU211002 MaxxTape, Ref: XTAPE; Lot BDDWTU211002 4) UDI (1 unit package): (01)00813954022079(17)250131(10)BDDWTU211002; (5 units package): (01)10813954022076(17)250131(10)BDDWTU211002 NEOTAPE, Lot BDDWTU211002 5) UDI (1 unit package): (01)00842517100084(17)250131(10)BDDWTU211002; (10 units package): (01)00842517100077(17)250131(10)BDDWTU211002 Karl Schumacher Dental, REF RCWT0001; Lot BDDWTU211002
510(k) numbers
["K152600"]
PMA numbers
not on file
Product code
KGN
Firm FEI
3008817245
Firm city
Oakland
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register