Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Navigation, Inc.: Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures.

Recall number
Z-1776-2024
Recalling firm
Medtronic Navigation, Inc.
Product
Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures.
Status
Open, Classified
Initiated
2024-03-26
Posted
2024-05-09
Terminated
not on file
Reason
Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.
Root cause
Software design
Action
On April 10, 2024, Medtronic Navigations, Inc. issued an Urgent: Medical Device Correction to affected consignees via UPS. In addition to informing consignees about the issue, Medtronic asked consignees to take the following actions: 1. Please review this information with all physician users. If you have any questions related to this issue, please contact Medtronic Technical Services for help at 1-888-826-5603 or email at rs.navtechsupport@medtronic.com. 2. Please confirm via the enclosed confirmation form that this notification has been communicated within your facility with all physician users. Send the completed Customer Confirmation Form to Medtronic via email at neuro.quality@medtronic.com. a. Medtronic will work expeditiously with your facility to coordinate a software update at no charge to resolve this anomaly. 3. Pass this notice on to those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. 4. Adverse reactions or quality problems experienced with this product should be reported to FDA and Medtronic. Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail). Call FDA at (800) FDA-1088. Call Medtronic Technical Services at 1-888-826-5603.
Quantity
3286 applications.
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Republic Of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, Nicaragua, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Lot, serial or model codes
UDIs: 00763000234867, 00763000125295, 00763000063177, 00643169865341/ Software Versions: 1.2.0, 1.1.0, 1.0.3, 1.0.2, and 1.0.1.
510(k) numbers
["K162309"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000517638
Firm city
Lafayette
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register