Skip to the content

FDA Medical Device Recalls (openFDA)

Volcano Corporation: Refinity Short Tip Rotational Intravascular Ultrasound (IVUS) Catheter; Product number: 400-0200.273; Catalog number 89900. Cardiovascular: The REFINITY (or REFINITY ST) Rotational IVUS Catheter (REFINITY or REFINITY ST) is intended for the intravascular ultrasound examination of coronary arteries. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. REFINITY (or REFINITY ST) catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

Recall number
Z-1776-2017
Recalling firm
Volcano Corporation
Product
Refinity Short Tip Rotational Intravascular Ultrasound (IVUS) Catheter; Product number: 400-0200.273; Catalog number 89900. Cardiovascular: The REFINITY (or REFINITY ST) Rotational IVUS Catheter (REFINITY or REFINITY ST) is intended for the intravascular ultrasound examination of coronary arteries. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. REFINITY (or REFINITY ST) catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.
Status
Terminated
Initiated
2017-03-29
Posted
2017-04-10
Terminated
2017-06-21
Reason
High complaint rates of kinking near the junction of the proximal shaft to telescope section of the catheter. Kinking could result in disruption of normal clinical workflow or loss of imaging.
Root cause
Under Investigation by firm
Action
Recall letters dated March 28, 2017 will be hand delivered by Volcano reps. Reps will also pick up any unused catheters for return and obtain a signed proof of delivery for both customer and Volcano Rep.
Quantity
718 total devices
Distribution
Distribution to US and Japan
Lot, serial or model codes
Lot # 50069132, Expiry date 6/30/2018; Lot # 50083527, Expiry date 11/30/2018; Lot # 50083533, Expiry Date 11/30/2018.
510(k) numbers
["K160583"]
PMA numbers
not on file
Product code
OBJ
Firm FEI
2939520
Firm city
Rancho Cordova
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register