Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Boston Scientific Radial Jaw 4 Large Capacity w/Needle Biopsy Forceps UPN/Catalog Number: MOO513330 The Radial Jaw 4 Large Capacity Inner Pouch, 240cm UPN M00513330 is a single unit which is packaged into a Box of 20 units (UPN M00513331), and a Box of 80 units (UPN M00513334). Single packaged units of M00513330 are not sold individually.

Recall number
Z-1774-2010
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific Radial Jaw 4 Large Capacity w/Needle Biopsy Forceps UPN/Catalog Number: MOO513330 The Radial Jaw 4 Large Capacity Inner Pouch, 240cm UPN M00513330 is a single unit which is packaged into a Box of 20 units (UPN M00513331), and a Box of 80 units (UPN M00513334). Single packaged units of M00513330 are not sold individually.
Status
Terminated
Initiated
2010-05-06
Posted
2010-06-04
Terminated
2011-09-20
Reason
Clevis may detach from the coil preventing the device to open and/or close. Delays in the procedure may result.
Root cause
Process control
Action
Boston Scientific issued Recall notification letters via Federal Express Priority mail on May 6, 2010. The notification instructs the account to check their inventory, remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests the completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product from their inventory. In the letter, customers with questions are provided with a phone and email at the firm to contact.
Quantity
not on file
Distribution
Nationwide; Panama.
Lot, serial or model codes
Lot: 13244654
510(k) numbers
not on file
PMA numbers
not on file
Product code
FCL
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register