Skip to the content

FDA Medical Device Recalls (openFDA)

XTANT Medical Holdings, Inc: XTANT MEDICAL Xpress Pedicle Screw System Dual Lead 6.5 x 50 mm, Part Number X073-6555-DL.

Recall number
Z-1773-2026
Recalling firm
XTANT Medical Holdings, Inc
Product
XTANT MEDICAL Xpress Pedicle Screw System Dual Lead 6.5 x 50 mm, Part Number X073-6555-DL.
Status
Open, Classified
Initiated
2026-02-18
Posted
2026-04-03
Terminated
not on file
Reason
Units from one lot were mislabeled with an incorrect part number, identifying the device as a shorter screw length than it physically is.
Root cause
Under Investigation by firm
Action
Xtant Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignee on 2/18/2026 via FedEx. The notice explained the reason for the correction, risk to health, and requested the following: 1. Review all devices from Lot 4375-01, including devices in loose stock and devices staged in implant trays or caddies, and verify device identification using the laser-etched markings on each implant to confirm that devices are stored or staged in the appropriate location consistent with their size and configuration. 2. If any device retains non-conforming labeling, relabel the device and discard of the incorrect label and application of the corrected labeling provided by Xtant Medical. 3. No product return is required unless a device cannot be confidently verified or corrected on site; please contact Xtant Medical for further guidance in such cases. 4. Complete Correction Acknowledgement to confirm completion of the above actions. For questions, Rebecca Lennemann, SVP, Quality & Regulatory Affairs, at 1-406-924-5878, Monday through Friday, 8 a.m. to 5 p.m. (Mountain Time).
Quantity
2 units
Distribution
US: OR
Lot, serial or model codes
UDI/DI M697X0736555DL1, Lot Number 4375-01
510(k) numbers
["K152132"]
PMA numbers
not on file
Product code
MNI
Firm FEI
3005031160
Firm city
Belgrade
Firm state
MT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register