Skip to the content

FDA Medical Device Recalls (openFDA)

Cardiocommand Inc.: CardioCommand, Inc. Transesophageal Cardiac Pacing and Recording System, Model 7A, REF 0101 - Product Usage: Indications for Use: - Temporary left atrial pacing to treat bradycardia, hypotension or AV junctional rhythm - Acceleration of heart rate as an alternative to exercise or drugs during diagnostic cardiac studies such as echocardiography or radionuclide ventriculography. - Antitachycardia pacing for cardioversion of supraventricular tachycardia (SVT), including atrial flutter and re-entrant atrial or atrioventricular paroxysmal tachycardia. - Esophageal electrocardiography for differential diagnosis of complex atrial arrhythmia

Recall number
Z-1773-2019
Recalling firm
Cardiocommand Inc.
Product
CardioCommand, Inc. Transesophageal Cardiac Pacing and Recording System, Model 7A, REF 0101 - Product Usage: Indications for Use: - Temporary left atrial pacing to treat bradycardia, hypotension or AV junctional rhythm - Acceleration of heart rate as an alternative to exercise or drugs during diagnostic cardiac studies such as echocardiography or radionuclide ventriculography. - Antitachycardia pacing for cardioversion of supraventricular tachycardia (SVT), including atrial flutter and re-entrant atrial or atrioventricular paroxysmal tachycardia. - Esophageal electrocardiography for differential diagnosis of complex atrial arrhythmia
Status
Terminated
Initiated
2019-05-10
Posted
not on file
Terminated
2021-04-12
Reason
Possible corrosion of the battery contact(s).
Root cause
Process control
Action
The consignees were notified by letter on 05/10/2019. The notice requested the return of the unit.
Quantity
5 units
Distribution
US Nationwide distribution in the states of CT, VW, OH, IL, DE.
Lot, serial or model codes
Serial Numbers: 8059, 8060, 8022, 8063, 8211
510(k) numbers
not on file
PMA numbers
["P860008"]
Product code
LPA
Firm FEI
1927197
Firm city
Tampa
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register