Skip to the content

FDA Medical Device Recalls (openFDA)

Advanced Sterilization Products: Sterrad Cyclesure 24 Biological Indicator, P/N: 14324 The STERRAD CYCLESURE 24 Biological Indicator is intended to be used as a standard method for frequent monitoring of the STERRAD Sterilization System cycles.

Recall number
Z-1773-2012
Recalling firm
Advanced Sterilization Products
Product
Sterrad Cyclesure 24 Biological Indicator, P/N: 14324 The STERRAD CYCLESURE 24 Biological Indicator is intended to be used as a standard method for frequent monitoring of the STERRAD Sterilization System cycles.
Status
Terminated
Initiated
2012-02-17
Posted
2012-06-06
Terminated
2013-10-17
Reason
The recall was initiated by Advanced Sterilization Products because they recently determined that certain lots of raw materials that were rejected during incoming inspection acceptance test were used for the manufacture of a specific lot range of Sterrad Cyclesure 24 Biological Indicator.
Root cause
Process control
Action
Advanced Sterilization Products a Johnson & Johnson company sent an Urgent Product Recall letter dated March 16, 2012, to all affected customers. The letter provided the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to read the "Details on the Affected Devices/Description of the problem" section in the letter. Customers were instructed to immediately identify and set aside all product listed above in a manner that ensures product will not be used. The lot number and the part number are printed on the label placed around the ampoules. Customers were instructed to return any affected product in accordance with the "Product Return Instructions" section of the letter. Customers were instructed to complete the requested information on the enclosed postage paid business reply card and return it to Stericycle even if they do not have the affected products listed on the letter in stock. Customers with questions were instructed to contact Stericycle at (877) 492-4795. For questions regarding this recall call 949-453-6400.
Quantity
2,284 cases (137,040 Biological Indicators)
Distribution
Worldwide Distribution - USA (nationwide) and the countries of AR, AU, BE, BR, CA, EC, HK, JP, MX, and PR
Lot, serial or model codes
Lot# 214117, 216117, 217117, 227117, 227118, and 228117.
510(k) numbers
["K103222"]
PMA numbers
not on file
Product code
FRC
Firm FEI
3003662624
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register