Skip to the content

FDA Medical Device Recalls (openFDA)

EKOS Corporation: Infusion System. EKOS EndoWave Infusion System. EKOS EndoWave Infusion System Drug Delivery Catheter with Ultrasound Core. Labeled as Sterile, Single Use Only. Connector clip is CIC Connector Clip. The infusion system is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytes, into the peripheral vasculature.

Recall number
Z-1773-2008
Recalling firm
EKOS Corporation
Product
Infusion System. EKOS EndoWave Infusion System. EKOS EndoWave Infusion System Drug Delivery Catheter with Ultrasound Core. Labeled as Sterile, Single Use Only. Connector clip is CIC Connector Clip. The infusion system is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytes, into the peripheral vasculature.
Status
Terminated
Initiated
2008-04-01
Posted
2008-08-20
Terminated
2011-05-13
Reason
Catheter disconnection. There is a potential for the catheter to disconnect from the Catheter Interface Cable (CIC) pod which connects to the Control Unit.
Root cause
Other
Action
Firm representatives will visit and distribute connector clips and instructions for use to consignees beginning 04/01/2008 at no cost to the consignee.
Quantity
218 units
Distribution
Class II Recall - Nationwide
Lot, serial or model codes
CIC Connector Clip Part Number: 700-4000; Lot numbers: 080324024, 080324025, 080324026, 080324027, 080325002, 080325003. EndoWave System Control Unit S/N: PT3B-001, PT3B-002, PT3B-004, PT3B-006, PT3B-007, PT3B-008, PT3B-009, PT3B-010, PT3B-011, PT3B-012, PT3B-013, PT3B-014, PT3B-017, PT3B-018, PT3B-020, PT3B-021, PT3B-022, PT3B-024, PT3B-025, PT3B-027, PT3B-029, PT3B-030, PT3B-031, PT3B-032, PT3B-033, PT3B-034, PT3B-035, PT3B-037, PT3B-038, PT3B-040, PT3B-041, PT3B-042, PT3B-043, PT3B-044, PT3B-045, PT3B-046, PT3B-047, PT3B-049, PT3B-050, PT3B-051, PT3B-052, PT3B-057, PT3B-058, PT3B-059, PT3B-060, PT3B-062, PT3B-063, PT3B-064, PT3B-065, PT3B-066, PT3B-067, PT3B-068, PT3B-070, PT3B-071, PT3B-072, PT3B-073, PT3B-074, PT3B-075, PT3B-076, PT3B-078, PT3B-079, PT3B-080, PT3B-082, PT3B-084, PT3B-085, PT3B-087, PT3B-088, PT3B-089, PT3B-090, PT3B-091, PT3B-093, PT3B-094, PT3B-095, PT3B-096, PT3B-097, PT3B-098, PT3B-099, PT3B-100, PT3B-102, PT3B-103, PT3B-104, PT3B-108, PT3B-109, PT3B-110, PT3B-111, PT3B-112, PT3B-113, PT3B-114, PT3B-115, PT3B-118, PT3B-119, PT3B-121, PT3B-123, PT3B-124, PT3B-127, PT3B-128, PT3B-130, PT3B-131, PT3B-132, PT3B-133, PT3B-135, PT3B-137, PT3B-138, PT3B-139, PT3B-140, PT3B-141, PT3B-143, PT3B-145, PT3B-146, PT3B-147, PT3B-148, PT3B-150, PT3B-151, PT3B-152, PT3B-153, PT3B-154, PT3B-155, PT3B-156, PT3B-157, PT3B-158, PT3B-159, PT3B-163, PT3B-164, PT3B-165, PT3B-168, PT3B-169, PT3B-172, PT3B-173, PT3B-174, PT3B-175, PT3B-176, PT3B-177, PT3B-178, PT3B-180, PT3B-181, PT3B-183, PT3B-186, PT3B-188, PT3B-192, PT3B-193, PT3B-195, PT3B-196, PT3B-197, PT3B-199, PT3B-201, PT3B-202, PT3B-204, PT3B-205, PT3B-206, PT3B-207, PT3B-208, PT3B-210, PT3B-211, PT3B-212, PT3B-214, PT3B-215, PT3B-217, PT3B-218, PT3B-219, PT3B-221, PT3B-253, PT3B-254, PT3B-255, PT3B-256, PT3B-257, PT3B-258, PT3B-259, PT3B-260, PT3B-261, PT3B-262, PT3B-263, PT3B-264, PT3B-265, PT3B-266, PT3B-267, PT3B-268, PT3B-269, PT3B-270, PT3B-271, PT3B-273, PT3B-274, PT3B-275, PT3B-276, PT3B-280, PT3B-285, PT3B-286, PT3B-287, PT3B-288, PT3B-289, PT3B-291, PT3B-292, PT3B-294, PT3B-295, PT3B-297, PT3B-298, PT3B-308, PT3B-310, PT3B-311, PT3B-312, PT3B-313, PT3B-314, PT3B-315, PT3B-316, PT3B-319, PT3B-320, PT3B-321, PT3B-323, PT3B-326, PT3B-328, PT3B-329, PT3B-332, PT3B-333, PT3B-334, PT3B-335, PT3B-337, PT3B-338, PT3B-339, and PT3B-344.
510(k) numbers
["K072507"]
PMA numbers
not on file
Product code
KRA
Firm FEI
3001627457
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register