Skip to the content

FDA Medical Device Recalls (openFDA)

Cardiovascular Systems Inc: ViperWire Advance Peripheral Guide Wire with Flex Tip, Model No. VPR-GW-FT18. Sterilized with Ethylene Oxide. The guide wire is a smooth, stainless steel wire, with a silicone coating, and a radiopaque distal spring tip. The guide wire allows for proper positioning of the device crown within peripheral arteries and provides a center of rotation for the device drive shaft. The guide wire torquer is a small, plastic accessory, packaged with the guide wire, and provides a gripping surface for manipulating the guide wire, if desired.

Recall number
Z-1771-2017
Recalling firm
Cardiovascular Systems Inc
Product
ViperWire Advance Peripheral Guide Wire with Flex Tip, Model No. VPR-GW-FT18. Sterilized with Ethylene Oxide. The guide wire is a smooth, stainless steel wire, with a silicone coating, and a radiopaque distal spring tip. The guide wire allows for proper positioning of the device crown within peripheral arteries and provides a center of rotation for the device drive shaft. The guide wire torquer is a small, plastic accessory, packaged with the guide wire, and provides a gripping surface for manipulating the guide wire, if desired.
Status
Terminated
Initiated
2017-03-03
Posted
not on file
Terminated
2017-11-28
Reason
Cardiovascular Systems, Inc. is recalling one lot of ViperWire Advance with Flextip, model VPR-GW-FT18 because it may contain a 0.014" ViperWire Advance Flextip product instead of the 0.018" ViperWire Advance with Flextip.
Root cause
Under Investigation by firm
Action
Customers were sent an Urgent Medical Device Recall letter beginning March 03, 2017. The letter identified affected product, reason for recall, and asked for affected devices to be removed from use and return to CSI. Questions can be directed to CSI Customer Service at 877-274-0901, or your CSI Sales Representative.
Quantity
186 devices
Distribution
Nationwide distribution in the states of AR, AZ, CO, FL, GA, HI, IN, KS, KY, LA, MS. NC, NM, RI, TX, VA
Lot, serial or model codes
Lot 156032
510(k) numbers
["K151260"]
PMA numbers
not on file
Product code
MCW
Firm FEI
3004742232
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register