Skip to the content

FDA Medical Device Recalls (openFDA)

Fresenius Medical Care Holdings, Inc.: AquaC UNO H [120V, 60Hz, 1700VA, US-Version- a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-0970-1 Vivonic Part Number: G02000966-US

Recall number
Z-1770-2022
Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
AquaC UNO H [120V, 60Hz, 1700VA, US-Version- a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-0970-1 Vivonic Part Number: G02000966-US
Status
Open, Classified
Initiated
2022-06-21
Posted
2022-09-19
Terminated
not on file
Reason
Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
Root cause
Under Investigation by firm
Action
Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) issued Urgent Medical Deice Correction letter (FA-2022-20-C) on 06/21/22 via First Class Mail. Letter states reason for recall, health risk and action to take: A Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) Water Systems Service Specialist will be contacting you to schedule a correction of your portable Water Treatment system. Only a Certified AquaC UNO H System Technician or FMCRTG Water Systems Service Specialist may perform this service. If choosing to correct via your own Certified AquaC UNO H System Technician, you will need to contact FMCRTG prior to performing the correction. FMCRTG will provide the necessary instruction via a Change Notification. For questions or concerns, please contact FMCRTG Technical Support Services at 888-553-1109.
Quantity
6 units
Distribution
US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.
Lot, serial or model codes
UDI: (01)4251285200234(11)200422(21)0CHS3987 (01)4251285200234(11)200526(21)0CHS4108 (01)4251285200234(11)210623(21)1CHS5372 (01)4251285200234(11)210707(21)1CHS5405 (01)4251285200234(11)190305(21)9CHS2881 (01)4251285200234(11)190529(21)9CHS3090 Serial Numbers: 0CHS3987 0CHS4108 1CHS5372 1CHS5405 9CHS2881 9CHS3090
510(k) numbers
["K143617"]
PMA numbers
not on file
Product code
FIP
Firm FEI
3001451489
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register