FDA Medical Device Recalls (openFDA)
Fresenius Medical Care Holdings, Inc.: AquaC UNO H [120V, 60Hz, 1700VA, US-Version- a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-0970-1 Vivonic Part Number: G02000966-US
- Recall number
- Z-1770-2022
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Product
- AquaC UNO H [120V, 60Hz, 1700VA, US-Version- a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-0970-1 Vivonic Part Number: G02000966-US
- Status
- Open, Classified
- Initiated
- 2022-06-21
- Posted
- 2022-09-19
- Terminated
- not on file
- Reason
- Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
- Root cause
- Under Investigation by firm
- Action
- Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) issued Urgent Medical Deice Correction letter (FA-2022-20-C) on 06/21/22 via First Class Mail. Letter states reason for recall, health risk and action to take: A Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) Water Systems Service Specialist will be contacting you to schedule a correction of your portable Water Treatment system. Only a Certified AquaC UNO H System Technician or FMCRTG Water Systems Service Specialist may perform this service. If choosing to correct via your own Certified AquaC UNO H System Technician, you will need to contact FMCRTG prior to performing the correction. FMCRTG will provide the necessary instruction via a Change Notification. For questions or concerns, please contact FMCRTG Technical Support Services at 888-553-1109.
- Quantity
- 6 units
- Distribution
- US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.
- Lot, serial or model codes
- UDI: (01)4251285200234(11)200422(21)0CHS3987 (01)4251285200234(11)200526(21)0CHS4108 (01)4251285200234(11)210623(21)1CHS5372 (01)4251285200234(11)210707(21)1CHS5405 (01)4251285200234(11)190305(21)9CHS2881 (01)4251285200234(11)190529(21)9CHS3090 Serial Numbers: 0CHS3987 0CHS4108 1CHS5372 1CHS5405 9CHS2881 9CHS3090
- 510(k) numbers
- ["K143617"]
- PMA numbers
- not on file
- Product code
- FIP
- Firm FEI
- 3001451489
- Firm city
- Waltham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28