Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corporation: Terumo Advanced Perfusion System 1 Electronic Patient Gas System (EPGS), Electronic Gas Blender, Catalog #: 801188, UDI 00886799000588. Used in cardio-pulmonary bypass.

Recall number
Z-1769-2019
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1 Electronic Patient Gas System (EPGS), Electronic Gas Blender, Catalog #: 801188, UDI 00886799000588. Used in cardio-pulmonary bypass.
Status
Terminated
Initiated
2018-10-17
Posted
not on file
Terminated
2020-04-30
Reason
An intermittent failure which causes a blender initialization fault upon start-up or during an EPGS calibration attempt leads to the Central Control Monitor control sliders being disabled, such that the gas system can only be controlled by the local control knobs.
Root cause
Process control
Action
On October 17, 2018, the firm notified customers of the recall via Urgent Medical Device Recall letter. The letter alerted customers that the devices were assembled with stepper motors that did not meet supplier specifications. Customers were informed that in the event of a failure, the gas system flow and FiO2 control sliders on the Central Control Monitor would be disabled and the gas system could only be controlled by the local control knobs. Customers were asked to take the following actions: 1. Review this Medical Device Correction and assure that all users have received notice of this issue. 2. Confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. 3. Terumo CVS will contact users to schedule the field correction activities. Note: Terumo CVS recommends that users continue using Terumo System 1 while waiting for this correction. If you have any questions, please contact Terumo CVS Customer Service: 1.800.521.2818, Monday Friday, 8 a.m. 6 p.m. ET Recall Fax: 1.734.741.6149
Quantity
72
Distribution
Distribution US nationwide, including CA, CT, IL, IN, KY, MI, MN, MO, MT, NM, NY, OH, and WI. Foreign distribution to Canada, Mexico, and South Korea.
Lot, serial or model codes
01338 01339 01340 01341 01342 01349 01350 01351 01352 01353 01354 01355 01356 01357 01358 01360 01361 01362 01363 01365 01366 01367 01369 01370 01371 01372 01373 01374 01375 01376 01377 01378 01379 Lot/Serial 01380 01381 01382 01383 01384 01385 01386 01387 01388 01389 01394 01395 01396 01397 01398 01405 01406 01407 01408 01410 01411 01413 01414 01415 01416 01418 01420 01421 01422 01423 01424 01425 01426 Lot/Serial 01427 01428 01429 01430 01431 01432 01434 01435 1718 2317 4417 9999  UDI 00886799000588
510(k) numbers
["K172220"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register